# Commission Decision of 7 May 2002 on common technical specifications for in vitro-diagnostic medical devices (notified under document number C(2002) 1344) (Text with EEA relevance) (2002/364/EC), Division 3: COMMON TECHNICAL SPECIFICATIONS (CTS) FOR PRODUCTS REFERRED TO IN ANNEX II, LIST A OF DIRECTIVE 98/79/EC

Extent: E+W+S+N.I.

Table 1
Screening assays: anti-HIV 1/2, HIV 1/2 Ag/Ab, anti-HTLV I/II, anti-HCV, HCV Ag/Ab, HBsAg, anti-HBc
anti-HIV 1/2, HIV 1/2 Ag/Ab
Anti-HTLV-I/II
anti-HCV, HCV Ag/Ab
HBsAg
Anti-HBc
Diagnostic sensitivity
Positive specimens
400 HIV-1
100 HIV-2
including 40 non-B-subtypes, all available HIV/1 subtypes shall be represented by at least 3 samples per subtype
300 HTLV-I
100 HTLV-II
400 (positive samples)
Including samples from different stages of infection and reflecting different antibody patterns.
Genotype 1-4: > 20 samples per genotype (including non-a subtypes of genotype 4);
5: > 5 samples;
6: if available
400
including subtype-consideration
400
including evaluation of other HBV-markers
Sero-conversion panels
20 panels
10 further panels (at Approved Body or manufacturer)
To be defined when available
20 panels
10 further panels (at Approved Body or manufacturer)
20 panels
10 further panels (at Approved Body or manufacturer)
To be defined when available
Analytical sensitivity
Standards
0,130 IU/ml (WHO International Standard: Third International Standard for HBsAg, subtypes ayw1/adw2, HBV genotype B4, NIBSC code: 12/226)
Specificity
Unselected donors (including first-time donors)
5 000
5 000
5 000
5 000
5 000
Hospitalized patients
200
200
200
200
200
Potentially cross-reacting blood-specimens (RF+, related viruses, pregnant women, etc.)
100
100
100
100
100
Table 2
NAT assays for HIV1, HCV, HBV, HTLV I/II (qualitative and quantitative; not molecular typing)
European Pharmacopoeia guideline.
Notes: Acceptance criteria for ‘whole system failure rate leading to false-neg results’ is 99/100 assays positive.
For quantitative NATs a study shall be performed on at least 100 positive specimens reflecting the routine conditions of users (e.g. no pre-selection of specimens). Comparative results with another NAT test system shall be generated in parallel.
For qualitative NATs a study on diagnostic sensitivity shall be performed using at least 10 sero-conversion panels. Comparative results with another NAT test system shall be generated in parallel.
HIV1
HCV
HBV
HTLV I/II
Acceptance criteria
NAT
qualitative
quantitative
qualitative
quantitative
qualitative
quantitative
qualitative
quantitative
As for HIV quantitative
As for HIV quantitative
As for HIV quantitative
Sensitivity
Detection limit
Detection of analytical sensitivity (IU/ml; defined on WHO standards or calibrated reference materials)
According to EP validation guideline : several dilution series into borderline concentration; statistical analysis (e.g. Probit analysis) on the basis of at least 24 replicates; calculation of 95 % cut-off value
Detection limit: as for qualitative tests; Quantification limit: dilutions (half-log10 or less) of calibrated reference preparations, definition of lower, upper quantification limit, precision, accuracy, ‘ linear ’ measuring range, ‘ dynamic range ’ . Reproducibility at different concentration levels to be shown
According to EP validation guideline : several dilution series into borderline concentration; statistical analysis (e.g. Probit analysis) on the basis of at least 24 replicates; calculation of 95 % cut-off value
According to EP validation guideline : several dilution series into borderline concentration; statistical analysis (e.g. Probit analysis) on the basis of at least 24 replicates; calculation of 95 % cut-off value
According to EP validation guideline : several dilution series into borderline concentration; statistical analysis (e.g. Probit analysis) on the basis of at least 24 replicates; calculation of 95 % cut-off value
Genotype/subtype detection/quantification efficiency
At least 10 samples per subtype (as far as available)
Dilution series of all relevant genotypes/subtypes, preferably of reference materials, as far as available
At least 10 samples per genotype (as far as available)
As far as calibrated genotype reference materials are available
As far as calibrated genotype reference materials are available
Cell culture supernatants (could substitute for rare HIV-1 subtypes)
Transcripts or plasmids quantified by appropriate methods may be used.
According to EP validation guideline as far as calibrated subtype reference materials are available; in vitro transcripts could be an option
According to EP validation guideline as far as calibrated subtype reference materials are available; in vitro transcripts could be an option
According to EP validation guideline as far as calibrated subtype reference materials are available; in vitro transcripts could be an option
According to EP validation guideline as far as calibrated subtype reference materials are available; in vitro transcripts could be an option
Diagnostic specificity negative samples
500 blood donors
100 blood donors
500 blood donors
500 blood donors
500 individual blood donations
Potential cross-reactive markers
By suitable assay design evidence (e.g. sequence comparison) and/or testing of at least 10 human retrovirus (e.g. HTLV)-positive samples
As for qualitative tests
By assays design and/or testing of at least 10 human flavivirus (e.g. HGV, YFV) positive samples
By assays design and/or testing of at least 10 other DNA-virus positive samples
By assay design and/or testing of at least 10 human retrovirus (e.g. HIV-) positive samples
Robustness
As for qualitative tests
Cross-contamination
At least 5 runs using alternating high positive (known to occur naturally) and negative samples
At least 5 runs using alternating high positive (known to occur naturally) and negative samples
At least 5 runs using alternating high positive (known to occur naturally) and negative samples
At least 5 runs using alternating high positive (known to occur naturally) and negative samples
Inhibition
Internal control preferably to go through the whole NAT procedure
Internal control preferably to go through the whole NAT procedure
Internal control preferably to go through the whole NAT procedure
Internal control preferably to go through the whole NAT procedure
Whole system failure rate leading to false-neg results
At least 100 samples virus-spiked with 3 × the 95 % pos cut-off concentration
At least 100 samples virus-spiked with 3 × the 95 % pos cut-off concentration
At least 100 samples virus-spiked with 3 × the 95 % pos cut-off concentration
At least 100 samples virus-spiked with 3 × the 95 % pos cut-off concentration
99/100 assays positive
Table 3
Rapid tests: anti-HIV 1 and 2, anti-HCV, HBsAg, anti-HBc, anti-HTLV I and II
Anti-HIV 1/2
Anti-HCV
HBsAg
Anti-HBc
Anti-HTLV I/II
Acceptance criteria
Diagnostic sensitivity
Positive specimens
Same criteria as for screening assays
Same criteria as for screening assays
Same criteria as for screening assays
Same criteria as for screening assays
Same criteria as for screening assays
Same criteria as for screening assays
Sero-conversion panels
Same criteria as for screening assays
Same criteria as for screening assays
Same criteria as for screening assays
Same criteria as for screening assays
Same criteria as for screening assays
Same criteria as for screening assays
Diagnostic specificity
Negative specimens
1 000 blood donations
1 000 blood donations
1 000 blood donations
1 000 blood donations
1 000 blood donations
≥ 99 % (anti-HBc: ≥ 96 %)
200 clinical specimens
200 clinical specimens
200 clinical specimens
200 clinical specimens
200 clinical specimens
200 samples from pregnant women
200 samples from pregnant women
200 samples from pregnant women
200 samples from pregnant women
100 potentially interfering samples
100 potentially interfering samples
100 potentially interfering samples
100 potentially interfering samples
100 potentially interfering samples
Table 4
Confirmatory/supplementary assays for anti-HIV 1 and 2, anti-HTLV I and II, anti-HCV, HBsAg
Acceptance criteria no neutralisation for HBsAg confirmatory assay.
Anti-HIV confirmatory assay
Anti-HTLV confirmatory assay
HCV supplementary assay
HBsAg confirmatory assay
Acceptance criteria
Diagnostic sensitivity
Positive specimens
200 HIV-1 and 100 HIV-2
200 HTLV-I and 100 HTLV-II
300 HCV (positive samples)
300 HBsAg
Correct identification as positive (or indeterminate), not negative
Including samples from different stages of infection and reflecting different antibody patterns
Including samples from different stages of infection and reflecting different antibody patterns.
Genotypes 1 – 4: > 20 samples (including non-a subtypes of genotype 4);
5: > 5 samples;
6: if available
Including samples from different stages of infection
20 ‘high pos’ samples (> 26 IU/ml); 20 samples in the cut-off range
Sero-conversion panels
15 sero-conversion panels/low titre panels
15 sero-conversion panels/low titre panels
15 sero-conversion panels/low titre panels
Analytical sensitivity
Standards
Second International Standard for HBsAg, subtype adw2, genotype A, NIBSC code: 00/588
Diagnostic specificity
Negative specimens
200 blood donations
200 blood donation
200 blood donations
10 false positives as available from the performance evaluation of the screening assay .
No false-positive results/ no neutralisation
200 clinical samples including pregnant women
200 clinical samples including pregnant women
200 clinical samples including pregnant women
50 potentially interfering samples, including samples with indeterminate results in other confirmatory assays
50 potentially interfering samples including samples with indeterminate results in other confirmatory assays
50 potentially interfering samples including samples with indeterminate results in other supplementary assays
50 potentially interfering samples
Table 5
HIV 1 antigen, HIV Ag/Ab, HCV antigen, HCV Ag/Ab
The total number of seroconversion panels for combined Ag/Ab assays (from tables 1 and 5) need not be greater than 30.
HIV-1 antigen and HIV Ag/Ab assays
HCV antigen and HCV Ag/Ab assays
Acceptance criteria
Diagnostic sensitivity
Positive specimens
50 HIV-1 antigen positive
50 cell culture supernatants including different HIV-1 subtypes and HIV-2
25 HCV core antigen and/or HCV RNA positive but anti-HCV negative samples, comprising HCV genotypes 1-6 (if a genotype is not available, a justification shall be made)
See general principle in § 3.1.8
Sero-conversion panels
20 sero-conversion panels/low titre panels
20 sero-conversion panels/low titre panels
Analytical sensitivity
Standards
HIV-1 p24 Antigen, First International Reference Reagent, NIBSC code: 90/636
HCV core antigen detection limit shall be investigated using dilutions of the WHO International HCV core antigen Standard: (HCV core Ag product code: PEI 129096/12)
For HIV-1 p24 antigen: ≤ 2 IU/ml
Diagnostic specificity
200 blood donations
200 clinical samples
50 potentially interfering samples
200 blood donations, 200 clinical samples, 50 potentially interfering samples
≥ 99,5 % after neutralisation or, if no neutralisation test available, after resolution of the sample status according to general principles in § 3.1.5
Table 6
Serotyping and genotyping assay: HCV
HCV serotyping and genotyping assay
Acceptance criteria
Diagnostic sensitivity
Positive specimens
200 (positive samples)
Including samples from different stages of infection and reflecting different antibody patterns.
Genotypes 1 – 4: > 20 samples (including non-a subtypes of genotype 4);
5: > 5 samples;
6: if available
≥ 95 % agreement between serotyping and genotyping
> 95 % agreement between genotyping and sequencing
Diagnostic specificity
Negative specimens
100
Table 7
HBV markers: anti-HBs, anti HBc IgM, anti-HBe, HBeAg
Anti-HBs
Anti-HBc IgM
Anti-HBe
HBeAg
Acceptance criteria
Diagnostic sensitivity
Positive specimens
100 vaccinees
200
200
200
≥ 98 %
100 naturally infected persons
Including samples from different stages of infection (acute/chronic, etc.)
The acceptance criteria should only be applied on samples from acute infection stage.
Including samples from different stages of infection (acute/chronic, etc.)
Including samples from different stages of infection (acute/chronic, etc.)
Sero-conversion panels
10 follow-ups or anti-HBs sero-conversions
When available
Analytical sensitivity
Standards
WHO First International Reference Preparation 1977; NIBSC, United Kingdom
HBe — Referenzantigen 82; PEI Germany
Anti-HBs: < 10 mIU/ml
Diagnostic specificity
Negative specimens
500
200 blood donations
200 blood donation
200 blood donations
≥ 98 %
Including clinical samples
200 clinical samples
200 clinical samples
200 clinical samples
50 potentially interfering samples
50 potentially interfering samples
50 potentially interfering samples
50 potentially interfering samples
Table 8
HDV markers: anti-HDV, anti-HDV IgM, delta antigen
Anti-HDV
Anti-HDV IgM
Delta antigen
Acceptance criteria
Diagnostic sensitivity
Positive specimens
100
50
10
≥ 98 %
Specifying HBV markers
Specifying HBV markers
Specifying HBV markers
Diagnostic specificity
Negative specimens
200
200
200
≥ 98 %
Including clinical samples
Including clinical samples
Including clinical samples
50 potentially interfering samples
50 potentially interfering samples
50 potentially interfering samples
Table 9
Blood group antigens in the ABO, Rh and Kell blood group systems
1
2
3
Specificity
Number of tests per recommended method
Total number of samples to be tested for a launch product
Total number of samples to be tested for a new formulation, or use of well-characterised reagents
Anti-ABO1 (anti-A), anti-ABO2 (anti-B), anti-ABO3 (anti-A,B)
500
3 000
1 000
Anti-RH1 (anti-D)
500
3 000
1 000
Anti-RH2 (anti-C), anti-RH4 (anti-c), anti-RH3 (anti-E)
100
1 000
200
Anti-RH5 (anti-e)
100
500
200
Anti-KEL1 (anti-K)
100
500
200
Table 11
Variant Creutzfeldt-Jakob disease (vCJD) assays for blood screening
Material
Number of specimens
Acceptance Criteria
Analytical sensitivity
vCJD brain spikes in human plasma (WHO reference number NHBY0/0003)
24 replicates of each of three dilutions of the material WHO number NHBY0/0003
(1×10 4 , 1×10 5 , 1×10 6 )
23 of the 24 replicates detected at
1×10 4
vCJD spleen spikes in human plasma (10 % spleen homogenate — NIBSC reference number NHSY0/0009)
24 replicates of each of three dilutions of the material NIBSC number NHSY0/0009
(1×10, 1×10 2 , 1×10 3 )
23 of the 24 replicates detected at
1×10
Diagnostic sensitivity
A)
Specimen from appropriate animal models
As many specimen as reasonably possible and available, and at least 10 specimens
90 %
B)
Specimen from humans with known clinical vCJD
As many specimen as reasonably possible and available, and at least 10 specimens
90 %
Only in case where 10 specimens are not available:
the number of specimens tested shall be comprised between 6 and 9
all available specimens shall be tested
no more than one false negative result
Analytical specificity
Potentially cross-reacting blood-specimens
100
Diagnostic specificity
Normal human plasma samples from area of low BSE exposure
5 000
At least 99,5 %

---