Division Acceptance criteria:F1
United Kingdom
All of the above reagents shall show comparable test results with established reagents with acceptable performance with regard to claimed reactivity of the device. For established reagents, where the application or use has been changed or extended, further testing should be carried out in accordance with the requirements outlined in column 1 (above).
Performance evaluation of anti-D reagents shall include tests against a range of weak RH1 (D) and partial RH1 (D) samples, depending on the intended use of the product.
Annotations
Textual amendments
Notes on Division 3
Textual amendments
- F2
- F3
Words in Annex Table 1 substituted (E.W.S.) (11.8.2021) by The Medical Devices (Amendment) (EU Exit) Regulations 2021 (S.I. 2021/873), reg. 1(1), Sch. 2 para. 3(d)
- F4
Editorial notes
Footnotes
- f1
European Pharmacopoeia guideline.
- f2
Acceptance criteria no neutralisation for HBsAg confirmatory assay.
- f3
The total number of seroconversion panels for combined Ag/Ab assays (from tables 1 and 5) need not be greater than 30.