Only by recommended techniques where reactivity against these antigens is claimed.

Note: Polyclonal reagents must be tested against a wider panel of cells to confirm specificity and exclude presence of unwanted contaminating antibodies.

Blood group reagents Minimum number of control cells to be tested
Positive reactions Negative reactions
A1 A2B Ax B 0
Anti-ABO1 (anti-A) 2 2 2 f1 2 2
B A1B A1 0
Anti-ABO2 (anti-B) 2 2 2 2
A1 A2 Ax B 0
Anti-ABO3 (anti-A,B) 2 2 2 2 4
R1r R2r WeakD r’r r’r rr
Anti-RH1 (anti-D) 2 2 2 1 1 1
R1R2 R1r r’r R2R2 r’r rr
Anti-RH2 (anti-C) 2 1 1 1 1 1
R1R2 R1r r’r R1R1
Anti-RH4 (anti-c) 1 2 1 3
R1R2 R2r r’r R1R1 r’r rr
Anti-RH 3 (anti-E) 2 1 1 1 1 1
R1R2 R2r r’r R2R2
Anti-RH5 (anti-e) 2 1 1 3
Kk kk
Anti-KEL1 (anti-K) 4 3
  1. Acceptance criteria:

Annotations

Textual amendments

F1

Substituted by Commission Decision of 27 November 2009 amending Decision 2002/364/EC on common technical specifications for in vitro diagnostic medical devices (notified under document C(2009) 9464) (Text with EEA relevance) (2009/886/EC).

Footnotes

f1

Only by recommended techniques where reactivity against these antigens is claimed.

Notes on Division 3

Textual amendments

F2

Substituted by Commission Implementing Decision (EU) 2019/1244 of 1 July 2019 amending Decision 2002/364/EC as regards requirements for HIV and HCV antigen and antibody combined tests and as regards requirements for nucleic acid amplification techniques with respect to reference materials and qualitative HIV assays (notified under document C(2019) 4632) (Text with EEA relevance).

F3

Words in Annex Table 1 substituted (E.W.S.) (11.8.2021) by The Medical Devices (Amendment) (EU Exit) Regulations 2021 (S.I. 2021/873), reg. 1(1), Sch. 2 para. 3(d)

F4

Inserted by Commission Decision of 20 December 2011 amending Decision 2002/364/EC on common technical specifications for in vitro diagnostic medical devices (notified under document C(2011) 9398) (Text with EEA relevance) (2011/869/EU).

Editorial notes

X1

Substituted by Corrigendum to Commission Decision 2009/886/EC of 27 November 2009 amending Decision 2002/364/EC on common technical specifications for in vitro diagnostic medical devices (Official Journal of the European Union L 318 of 4 December 2009).

Footnotes

f1

European Pharmacopoeia guideline.

f2

Acceptance criteria no neutralisation for HBsAg confirmatory assay.

f3

The total number of seroconversion panels for combined Ag/Ab assays (from tables 1 and 5) need not be greater than 30.