Division 3.1.2 F1
United Kingdom
Devices intended by the manufacturer for testing body fluids other than serum or plasma, e.g. urine, saliva, etc., shall meet the same CTS requirements for sensitivity and specificity as serum or plasma tests. The performance evaluation shall test samples from the same individuals in both the tests to be approved and in a respective serum or plasma assay.
Annotations
Textual amendments
Notes on Division 3
Textual amendments
- F2
- F3
Words in Annex Table 1 substituted (E.W.S.) (11.8.2021) by The Medical Devices (Amendment) (EU Exit) Regulations 2021 (S.I. 2021/873), reg. 1(1), Sch. 2 para. 3(d)
- F4
Editorial notes
Footnotes
- f1
European Pharmacopoeia guideline.
- f2
Acceptance criteria no neutralisation for HBsAg confirmatory assay.
- f3
The total number of seroconversion panels for combined Ag/Ab assays (from tables 1 and 5) need not be greater than 30.