# Commission Decision of 7 May 2002 on common technical specifications for in vitro-diagnostic medical devices (notified under document number C(2002) 1344) (Text with EEA relevance) (2002/364/EC), Division 3.1.5

Extent: E+W+S+N.I.

If discrepant test results are identified as part of an evaluation, these results shall be resolved as far as possible, for example:
by evaluation of the discrepant sample in further test systems,
by use of an alternative method or marker,
by a review of the clinical status and diagnosis of the patient, and
by the testing of follow-up-samples.

---