Division 3.1.8 F1
United Kingdom
Sensitivity with true positives and sero-conversion samples shall be evaluated as follows:
- 3.1.8.1.
Diagnostic test sensitivity during sero-conversion has to represent the state of the art. Whether further testing of the same or additional sero-conversion panels is conducted by the F2approved body or by the manufacturer the results shall confirm the initial performance evaluation data (see Table 1). Sero-conversion panels should start with a negative bleed(s) and should have narrow bleeding intervals.
- 3.1.8.2.
For blood screening devices (with the exception of HBsAg and anti-HBc tests), all true positive samples shall be identified as positive by the device to be F3UK or CE marked, (Table 1). For HBsAg and anti-HBc tests the new device shall have an overall performance at least equivalent to that of the established device (see 3.1.4).
- 3.1.8.3.
Regarding HIV tests:
all sero-conversion HIV samples shall be identified as positive, and
at least 40 early sero-conversion HIV samples shall be tested. Results should conform to the state of the art.
Annotations
Textual amendments
- F1
- F2
Words in Annex point 3.1.8 substituted (E.W.S.) (11.8.2021) by The Medical Devices (Amendment) (EU Exit) Regulations 2021 (S.I. 2021/873), reg. 1(1), Sch. 2 para. 3(b)(ii)
- F3
Words in Annex point 3.1.8 substituted (E.W.S.) (11.8.2021) by The Medical Devices (Amendment) (EU Exit) Regulations 2021 (S.I. 2021/873), reg. 1(1), Sch. 2 para. 3(b)(i)
Notes on Division 3
Textual amendments
- F2
- F3
Words in Annex Table 1 substituted (E.W.S.) (11.8.2021) by The Medical Devices (Amendment) (EU Exit) Regulations 2021 (S.I. 2021/873), reg. 1(1), Sch. 2 para. 3(d)
- F4
Editorial notes
Footnotes
- f1
European Pharmacopoeia guideline.
- f2
Acceptance criteria no neutralisation for HBsAg confirmatory assay.
- f3
The total number of seroconversion panels for combined Ag/Ab assays (from tables 1 and 5) need not be greater than 30.