Division 3.2.2 F1F2
United Kingdom
Hepatitis C virus (HCV) antigen and antibody combined tests intended for the detection of HCV antigen and HCV antibody shall meet the requirements for sensitivity and specificity set out in Table 1 and Table 5. HCV seroconversion panels for the evaluation of HCV antigen and antibody combined tests shall start with one or more negative bleeds and comprise panel members from early HCV infection (HCV core antigen and/or HCV RNA positive but anti-HCV negative). HCV antigen and antibody combined tests shall demonstrate enhanced sensitivity in early HCV infection when compared to HCV antibody only tests.
Annotations
Textual amendments
- F1
- F2
Notes on Division 3
Textual amendments
- F3
Words in Annex Table 1 substituted (E.W.S.) (11.8.2021) by The Medical Devices (Amendment) (EU Exit) Regulations 2021 (S.I. 2021/873), reg. 1(1), Sch. 2 para. 3(d)
- F4
Editorial notes
Footnotes
- f1
European Pharmacopoeia guideline.
- f2
Acceptance criteria no neutralisation for HBsAg confirmatory assay.
- f3
The total number of seroconversion panels for combined Ag/Ab assays (from tables 1 and 5) need not be greater than 30.