In the case of repeated analyses of a certified reference material, the guideline ranges for the deviation of the experimentally determined recovery corrected mean mass fraction from the certified value are as follows:

Table 2

Minimum trueness of quantitative methods

Mass fractionRange
≤ 1 μg/kg– 50 % to + 20 %
> 1 μg/kg to 10 μg/kg– 30 % to + 10 %
≥ 10 μg/kg– 20 % to + 10 %

When no such CRMs are available, it is acceptable that trueness of measurements is assessed through recovery of additions of known amounts of the analyte(s) to a blank matrix. Data corrected with the mean recovery are only acceptable when they fall within the ranges shown in Table 2.