In the case of repeated analysis of a sample carried out under within-laboratory reproducibility conditions, the intra-laboratory coefficient of variation (CV) of the mean shall not exceed the following values:

Table 8

CVs for quantitative methods at a range of element mass fractions

Mass fractionCV (%)
≥ 10 μg/kg to 100 μg/kg20
> 100 μg/kg to 1 000 μg/kg15
≥ 1 000 μg/kg10