The Annex to Decision 2002/364/EC is amended as follows:

  1. 1..

    Section 2 is amended as follows:

    1. (a)

      the following definition of ‘First line assay’ is inserted between the definition of ‘Whole system failure rate’ and the definition of ‘Confirmation assay’:

      ‘ First-line assay

      First-line assay means an assay used to detect a marker or analyte, and which may be followed by a confirmatory assay. Devices intended solely to be used to monitor a previously determined marker or analyte are not considered first-line assays.’;

    2. (b)

      the definition of ‘Confirmation assay’ is replaced by the following:

      ‘ Confirmatory assay

      Confirmatory assay means an assay used for the confirmation of a reactive result from a first-line assay.’;

  2. 2..

    Section 3 is amended as follows:

    1. (a)

      Sub-section 3.1.1 is replaced by the following:

      1. 3.1.1.

        ‘Devices which detect virus infections shall meet the requirements for sensitivity and specificity set out in Table 1, Table 3, Table 4 and Table 5, which apply to them taking account of the intended purpose of the devices concerned, virus type and entities to be detected (antigen and/or antibody). See also principle 3.1.11 for first-line assays.’;

    2. (b)

      Sub-section 3.1.3 is replaced by the following:

      1. 3.1.3.

        ‘Devices for self-testing shall meet the same CTS requirements for sensitivity and specificity as respective devices for professional use. Relevant parts of the performance evaluation shall be carried out (or repeated) by appropriate lay users to validate the operation of the device and the instructions for use. The lay users selected for the performance evaluation shall be representative of the intended users groups.

        Performance evaluation of a device for self-testing shall include, for each body fluid claimed for use with the device, e.g. whole blood, urine, saliva, etc., at least 200 lay users that are known positive for the infection and at least 400 lay users that do not know their status, of which at least 200 are at high risk of acquiring the infection. The sensitivity and specificity of the device for self-testing in the hands of lay users shall be defined against the confirmed patient infectious status.’;

    3. (c)

      Sub-section 3.1.9 is replaced by the following:

      1. 3.1.9.

        ‘Performance evaluation of first line assays shall include 25 positive (if available in the case of rare infections) ‘same day’ fresh serum samples (≤ 1 day after sampling).’;

    4. (d)

      Sub-section 3.1.11 is replaced by the following:

      1. 3.1.11.

        ‘For performance evaluations for first line assays (Table 1 and Table 3) blood donor populations shall be investigated from at least two blood donation centers and consist of consecutive blood donations, which have not been selected to exclude first time donors.’;

    5. (e)

      Sub-section 3.4.2 is replaced by the following:

      1. 3.4.2.

        ‘The manufacturer’s batch release testing for first line assays shall include at least 100 specimens negative for the relevant analyte.’.

  3. 3..

    Table 1 is replaced by the following:

    ‘

    Table 1

    First-line assays, excluding rapid tests: anti-HIV 1/2, HIV 1/2 Ag/Ab, anti-HTLV I/II, anti-HCV, HCV Ag/Ab, HBsAg, anti-HBc

    anti-HIV 1/2, HIV 1/2 Ag/AbAnti-HTLV-I/IIanti-HCV, HCV Ag/AbHBsAgAnti-HBc
    Diagnostic sensitivityPositive specimens

    400 HIV-1

    100 HIV-2

    including 40 non-B-subtypes, all available HIV/1 subtypes shall be represented by at least 3 samples per subtype

    300 HTLV-I

    100 HTLV-II

    400 (positive samples)

    Including samples from different stages of infection and reflecting different antibody patterns.

    Genotype 1-4: > 20 samples per genotype (including non-a subtypes of genotype 4);

    5: > 5 samples;

    6: if available

    400

    including subtype-consideration

    400

    including evaluation of other HBV-markers

    Sero-conversion panels

    20 panels

    10 further panels (at notified body or manufacturer)

    To be defined when available

    20 panels

    10 further panels (at notified body or manufacturer)

    20 panels

    10 further panels (at notified body or manufacturer)

    To be defined when available
    Analytical sensitivityStandards0,130 IU/ml ( WHO International Standard: Third International Standard for HBsAg, subtypes ayw1/adw2, HBV genotype B4, NIBSC code: 12/226)
    SpecificityUnselected donors (including first-time donors)5 0005 0005 0005 0005 000
    Hospitalized patients200200200200200
    Potentially cross-reacting blood-specimens (RF+, related viruses, pregnant women, etc)100100100100100

    ’

  4. 4..

    Table 3 is replaced by the following:

    ‘

    Table 3

    Rapid tests: anti-HIV 1/2, HIV 1/2 Ag/Ab, anti-HCV, HCV Ag/Ab, HBsAg, anti-HBc, anti-HTLV I and II

    anti-HIV 1/2, HIV 1/2 Ag/Abanti-HCV, HCV Ag/AbHBsAganti-HBcanti-HTLV I and IIAcceptance criteria
    Diagnostic sensitivityPositive specimensSame criteria as in Table 1Same criteria as in Table 1Same criteria as in Table 1Same criteria as in Table 1Same criteria as in Table 1Same criteria as in Table 1
    Sero-conversion panelsSame criteria as in Table 1Same criteria as in Table 1Same criteria as in Table 1Same criteria as in Table 1Same criteria as in Table 1Same criteria as in Table 1
    Diagnostic specificityNegative specimens 21 000 blood donations1 000 blood donations1 000 blood donations1 000 blood donations1 000 blood donations

    ≥ 99 %

    (anti-HBc: ≥ 96 %)

    200 clinical specimens200 clinical specimens200 clinical specimens200 clinical specimens200 clinical specimens
    200 samples from pregnant women200 samples from pregnant women200 samples from pregnant women200 samples from pregnant women
    100 potentially interfering samples100 potentially interfering samples100 potentially interfering samples100 potentially interfering samples100 potentially interfering samples

    ’

  5. 5..

    Table 4 is replaced by the following:

    ‘

    Table 4

    Confirmatory and supplementary assays for anti-HIV 1/ 2, HIV 1/2 Ag/Ab, anti-HTLV I and II, anti-HCV, HCV Ag/Ab, HBsAg

    Acceptance criteria: no neutralisation for HBsAg confirmatory assay.

    anti-HIV 1/2, HIV 1/2 Ag/Ab confirmatory assaysanti-HTLV I and II confirmatory assaysanti-HCV, HCV Ag/Ab supplementary assaysHBsAg confirmatory assaysAcceptance criteria
    Diagnostic sensitivityPositive specimens

    200 HIV-1 and 100 HIV-2

    Including samples from different stages of infection and reflecting different antibody patterns

    200 HTLV-I and 100 HTLV-II

    300 HCV (positive samples)

    Including samples from different stages of infection and reflecting different antibody patterns.

    Genotypes 1 – 4: > 20 samples (including non-a subtypes of genotype 4;

    Genotype 5: > 5 samples;

    Genotype 6: if available

    300 HBsAg

    Including samples from different stages of infection

    20 ‘high pos’ samples (>26 IU/ml); 20 samples in the cut-off range

    Correct identification as positive (or indeterminate), not negative
    Sero-conversion panels15 sero-conversion panels/low titre panels15 sero-conversion panels/low titre panels15 sero-conversion panels/low titre panels
    Analytical sensitivityStandardsThird International Standard for HBsAg, subtypes ayw1/adw2, HBV genotype B4, NIBSC code: 12/226
    Diagnostic specificityNegative specimens200 blood donations200 blood donations200 blood donations10 false positives as available from the performance evaluation of the first line assayf1.No false-positive results/ no neutralisation
    200 clinical samples including pregnant women200 clinical samples including pregnant women200 clinical samples including pregnant women
    50 potentially interfering samples, including samples with indeterminate results in other confirmatory assays50 potentially interfering samples, including samples with indeterminate results in other confirmatory assays50 potentially interfering samples, including samples with indeterminate results in other supplementary assays50 potentially interfering samples

    ’

Annotations

Footnotes

f1

Acceptance criteria: no neutralisation for HBsAg confirmatory assay.