OpenLegislation

Division 5.3.3.1 Column

United Kingdom

Original text. Later amendments are not shown.

This is in two parts, the first containing didecyl phthalate on Gas Chrom Q as packing, the second having Ucon 50 HB 280X on Gas Chrom Q as packing. The prepared combined column must yield a resolution ‘R’ equal to, or better than, 1,5, where:

R= 2 d' r 2 - r 1 W 1 + W 2

let:

r1 and r2
  • retention times (in minutes),

  • W1 and W2
  • peak widths at half height (in millimetres),

  • d'
  • the chart speed (in millimetres per minute).

  • As examples the following two parts yield the required resolution:

    Column A:

    Material: stainless steel.

    Length: 1,5 m.

    Diameter: 3 mm.

    Packing: 20 % didecyl phthalate on Gas Chrom Q (100 to 120 mesh).

    Column B:

    Materia: stainless steel.

    Length: 1,5 m.

    Diameter: 3 mm.

    Packing: 20 % Ucon 50 HB 280X on Gas Chrom Q (100 to 120 mesh).