Devices must be designed and manufactured in such a way as to remove or minimize as far as possible:
-
the risk of physical injury in connection with their physical, including dimensional, features,
-
risks connected with the use of energy sources with particular reference, where electricity is used, to insulation, leakage currents and overheating of the devices,
-
risks connected with reasonably foreseeable environmental conditions such as magnetic fields, external electrical influences, electrostatic discharge, pressure or variations in pressure and acceleration,
-
risks connected with medical treatment, in particular those resulting from the use of defibrillators or high-frequency surgical equipment,
-
F1risks connected with ionising radiation from radioactive substances included in the device, in compliance with the protection requirements laid down in Council Directive 96/29/Euratom of 13 May 1996 laying down basic safety standards for the protection of the health of workers and the general public against the dangers arising from ionising radiation f1 and Council Directive 97/43/Euratom of 30 June 1997 on health protection of individuals against the dangers of ionising radiation in relation to medical exposure f2 ,
-
risks which may arise where maintenance and calibration are impossible, including:
-
excessive increase of leakage currents,
-
ageing of the materials used,
-
excess heat generated by the device,
-
decreased accuracy of any measuring or control mechanism.
-