# Council Directive 93/42/EEC of 14 June 1993 concerning medical devices, Division 9.2

Extent: E+W+S+N.I.

Devices must be designed and manufactured in such a way as to remove or minimize as far as is possible:
the risk of injury, in connection with their physical features, including the volume/pressure ratio, dimensional and where appropriate ergonomic features,
risks connected with reasonably foreseeable environmental conditions, such as magnetic fields, external electrical influences, electrostatic discharge, pressure, temperature or variations in pressure and acceleration,
the risks of reciprocal interference with other devices normally used in the investigations or for the treatment given,
risks arising where maintenance or calibration are not possible (as with implants), from ageing of materials used or loss of accuracy of any measuring or control mechanism.

---