  "heading": "Rule 4",
  "text": "All non-invasive devices which come into contact with injured skin:\nare in Class I if they are intended to be used as a mechanical barrier, for compression or for absorption of exudates,\nare in Class IIb if they are intended to be used principally with wounds which have breached the dermis and can only heal by secondary intent,\nare in Class IIa in all other cases, including devices principally intended to manage the micro-environment of a wound.",