Division 2.2 Rule 6F1
United Kingdom
All surgically invasive devices intended for transient use are in Class IIa unless they are:
-
intended specifically to control, diagnose, monitor or correct a defect of the heart or of the central circulatory system through direct contact with these parts of the body, in which case they are in Class III,
-
reusable surgical instruments, in which case they are in Class I,
-
intended specifically for use in direct contact with the central nervous system, in which case they are in Class III,
-
intended to supply energy in the form of ionising radiation in which case they are in Class IIb,
-
intended to have a biological effect or to be wholly or mainly absorbed in which case they are in Class IIb,
-
intended to administer medicines by means of a delivery system, if this is done in a manner that is potentially hazardous taking account of the mode of application, in which case they are in Class IIb.