Division Preliminary noteF1
United Kingdom
In the table, ‘ Recommended ’ means that the measures should in principle be applied, unless the results of the assessment referred to in Article 3(2) indicate otherwise.
Group 1 biological agents
For work with group 1 biological agents including live attenuated vaccines, the principles of good occupational safety and hygiene should be observed.
Groups 2, 3 and 4 biological agents
It may be appropriate to select and combine containment requirements from different categories below on the basis of a risk assessment related to any particular process or part of a process.
| HEPA: High efficiency particulate air | |||
| Closed system: A system that physically separates the process from the environment (e.g. incubator vats, tanks, etc.). | |||
| Airlock: Entry must be through an airlock which is a chamber isolated from the laboratory. The clean side of the airlock must be separated from the restricted side by changing or showering facilities and preferably by interlocking doors. | |||
| A. Containment measures | B. Containment levels | ||
|---|---|---|---|
| 2 | 3 | 4 | |
| General | |||
| 1. Viable organisms should be handled in a system which physically separates the process from the environment | Yes | Yes | Yes |
| 2. Exhaust gases from the closed system should be treated so as to: | Minimise release | Prevent release | Prevent release |
| 3. Sample collection, addition of materials to a closed system and transfer of viable organisms to another closed system, should be performed so as to: | Minimise release | Prevent release | Prevent release |
| 4. Bulk culture fluids should not be removed from the closed system unless the viable organisms have been: | Inactivated by validated chemical or physical means | Inactivated by validated chemical or physical means | Inactivated by validated chemical or physical means |
| 5. Seals should be designed so as to: | Minimise release | Prevent release | Prevent release |
| 6. The controlled area should be designed to contain spillage of the entire contents of the closed system | No | Recommended | Yes |
| 7. The controlled area should be sealable to permit fumigation | No | Recommended | Yes |
| Facilities | |||
| 8. Decontamination and washing facilities should be provided for personnel | Yes | Yes | Yes |
| Equipment | |||
| 9. Input air and extract air to the controlled area should be HEPA f1 filtered | No | Recommended | Yes |
| 10. The controlled area should be maintained at an air pressure negative to atmosphere | No | Recommended | Yes |
| 11. The controlled area should be adequately ventilated to minimise air contamination | Recommended | Recommended | Yes |
| System of work | |||
| 12. Closed systems f2 should be located within a controlled area | Recommended | Recommended | Yes, and purpose-built |
| 13. Biohazard signs should be posted | Recommended | Yes | Yes |
| 14. Access should be restricted to nominated personnel only | Recommended | Yes | Yes, via an airlock f3 |
| 15. Personnel should shower before leaving the controlled area | No | Recommended | Yes |
| 16. Personnel should wear protective clothing | Yes, work clothing | Yes | Yes, complete change |
| Waste | |||
| 17. Effluent from sinks and showers should be collected and inactivated before release | No | Recommended | Yes |
| 18. Effluent treatment before final discharge | Inactivated by validated chemical or physical means | Inactivated by validated chemical or physical means | Inactivated by validated chemical or physical means |
Annotations
Textual amendments
Footnotes
- f1
-
HEPA: High efficiency particulate air
- f2
-
Closed system: A system that physically separates the process from the environment (e.g. incubator vats, tanks, etc.).
- f3
-
Airlock: Entry must be through an airlock which is a chamber isolated from the laboratory. The clean side of the airlock must be separated from the restricted side by changing or showering facilities and preferably by interlocking doors.