# Directive 2014/32/EU of the European Parliament and of the Council of 26 February 2014 on the harmonisation of the laws of the Member States relating to the making available on the market of measuring instruments (recast) (Text with EEA relevance), Division 1: Rated Operating Conditions

Extent: E+W+S+N.I.

The manufacturer shall specify the rated operating conditions for the instrument as follows:
1.1. For the measurand:
The measuring range in terms of its maximum and minimum capacity.
1.2. For the electrical supply influence quantities:
In case of AC voltage supply
:
the nominal AC voltage supply, or the AC voltage limits.
In case of DC voltage supply
:
the nominal and minimum DC voltage supply, or the DC voltage limits.
1.3. For the mechanical and climatic influence quantities:
The minimum temperature range is 30 °C unless specified otherwise in the following chapters of this Annex.
The mechanical environment classes according to Annex I, point 1.3.2 are not applicable. For instruments which are used under special mechanical strain, e.g. instruments incorporated into vehicles, the manufacturer shall define the mechanical conditions of use.
1.4. For other influence quantities (if applicable):
The rate(s) of operation.
The characteristics of the product(s) to be weighed.

---
Source: https://www.legislation.gov.uk/eudr/2014/32/annex/VIII/division/3/chapter/I/division/1 (the official version)
Page: https://openlegislation.uk/eu-directive-2014-32/annex-VIII/division-3/chapter-I/division-1
Contains public sector information licensed under the Open Government Licence v3.0. Source: legislation.gov.uk, The National Archives.