OpenLegislation

Appendix F1

United Kingdom

Original text. Later amendments are not shown.

Determination of the linear velocity of the gas

Inject 1-3 μl methane (or propane) into the GC apparatus after it has been regulated to normal operating conditions. Measure the time it takes for the gas to flow through the column from the time it is injected to the time the peak appears (t M ).

The linear velocity in cm/s is given by the formula L/t M , where L is the length of the column in cm and t M the time measured in seconds.

Annotations

Textual amendments

F1

Substituted by Commission Regulation (EC) No 702/2007 of 21 June 2007 amending Commission Regulation (EEC) No 2568/91 on the characteristics of olive oil and olive-residue oil and on the relevant methods of analysis.