Calibration method using a twin-pan balance:
A.4.1.2.1. Place the tare bottle on the left-hand pan and the clean, dry pycnometer with its collecting stopper on the right-hand pan. Balance them by placing weights on the pycnometer side: p grams.
A.4.1.2.2. Fill the pycnometer carefully with distilled water at ambient temperature and fit the thermometer; carefully wipe the pycnometer dry and place it in the thermally insulated jacket; agitate by inverting the container until the thermometer's temperature reading is constant.
Accurately adjust the level to the upper rim of the side tube. Clean the side tube, fit the collecting stopper; read the temperature t °C carefully and if necessary correct for any inaccuracies in the temperature scale.
Weigh the water-filled pycnometer, with p′ the weight in grams making up the equilibrium.
A.4.1.2.3.
Calculation
Tare of the empty pycnometer = p + m
where m is the mass of air in the pycnometer.
m = 0,0012 × (p – p′)
Volume of the pycnometer at 20 °C:
where Ft is the factor for temperature t °C taken from Table I of Chapter 1 ‘Density and specific gravity’ of the Annex to Regulation (EEC) No 2676/90 (p. 10).
V20 °C must be known to the nearest 0,001 ml.
Mass of water in the pycnometer at 20 °C:
M 20 °C = V 20 °C × 0,998203 where 0,998203 is the density of water at 20 °C.

---