Division 6 Apparatus and equipment
United Kingdom
- 6.1.
Apparatus capable of measuring the density and alcoholic strength.
- 6.2.
Analytical balance, capable of measuring to four decimal places.
- 6.3.
A temperature programmed gas chromatograph fitted with a flame ionisation detector and integrator or other data handling system capable of measuring peak areas or peak heights.
- 6.4.
Gas chromatographic column(s), capable of separating the analytes such that the minimum resolution between the individual components (other than 2-methylbutan-1-ol and 3-methylbutan-1-ol) is at least 1.3.
Note 2:
The following columns and GC conditions are suitable examples:
- 1..
A retention gap 1 m × 0,32 mm i.d. connected to a CP-WAX 57 CB column 50 m × 0,32 mm i.d. 0,2 μm film thickness (stabilised polyethylene glycol) followed by a Carbowax 400 column 50 m × 0,32 mm i.d. 0,2 μm film thickness. (Columns are connected using press-fit connectors.)
Carrier gas and pressure: Helium (135 kPa) Column temperature: 35 °C for 17 min., 35 to 70 °C at 12 °C/min., hold at 70 °C for 25 min. Injector temperature: 150 °C Detector temperature: 250 °C Injection volume: 1 μl, split 20 to 100:1 - 2..
A retention gap 1 m × 0,32 mm i.d. connected to a CP-WAX 57 CB column 50 m × 0,32 mm i.d. 0,2 μm film thickness (stabilised polyethylene glycol). (Retention gap is connected using a press-fit connector.)
Carrier gas and pressure: Helium (65 kPa) Column temperature: 35 °C for 10 min., 35 to 110 °C at 5 °C/min., 110 to 190 °C at 30 °C/min., hold at 190 °C for 2 min. Injector temperature: 260 °C Detector temperature: 300 °C Injection volume: 1 μl, split 55:1 - 3..
A packed column (5 % CW 20M, Carbopak B), 2 m × 2 mm i.d.
Column temperature: 65 °C for 4 min., 65 to 140 °C at 10 °C/min., hold at 140 °C for 5 min., 140 to 150 °C at 5 °C/min., hold at 150 °C for 3 min. Injector temperature: 65 °C Detector temperature: 200 °C Injection volume: 1 μl
- 1..