An automated system of data handling can be used, provided the data can be checked using the principles described in the method below.

Measure either peak areas or peak heights for congener and internal standard peaks.

  1. 9.1.

    Response factor calculation.

    From the chromatogram of the injection of standard solution C (5.14.3), calculate response factors for each congener using equation (1). where:

    IS
  2. Internal Standard

  3. Conc. congener
  4. concentration of congener in solution C (5.14.3)

  5. Conc. IS
  6. concentration of internal standard in solution C (5.14.3).

  1. 9.1.2.

    Sample analysis

    Using equation (2) below, calculate the concentration of each congener in the samples. where:

    MSAMPLE
  2. weight of sample (8.1.2);

  3. MIS
  4. weight of internal standard (8.1.3);

  5. Conc. IS
  6. concentration of internal standard in solution E (5.14.5);

  7. RF
  8. response factor calculated using equation 1.

  • 9.1.3.

    Quality control standard solution analysis

    Using equation (3) below, calculate the percentage recovery of the target value for each congener in the quality control standards (5.14.7):

    The concentration of the analyte in the QC standard is calculated using equations (1) and (2) above.

  • 9.2.

    Final presentation of results

    Results are converted from μ/g to g per 100 litres absolute alcohol for samples using equation (4): where

    ρ
  • density in kg/m3.

  • Results are quoted to 3 significant figures and a maximum of one decimal place e.g. 11,4 g per 100 l absolute alcohol.