Division 4.3.2.3.2 Initial validation through collaborative trialsF1
United Kingdom
Validation through collaborative trials shall be done in accordance with an internationally recognised protocol on collaborative trials (e.g. ISO 5725:1994 or the IUPAC International Harmonised Protocol) which requires inclusion of valid data from at least eight different laboratories. Other than that, the only difference compared to single laboratory validations is that the ≥ 20 samples per commodity/level can be evenly divided over the participating laboratories, with a minimum of two samples per laboratory.