For in-house validated methods, as an alternative a ‘fitness-for-purpose’ approach may be used to assess their suitability for official control. Methods suitable for official control must produce results with a combined standard measurement uncertainty (u) less than the maximum standard measurement uncertainty calculated using the formula below:
Uf=LOD∕22+αC2
where:
Uf is the maximum standard measurement uncertainty (μg/kg).
LOD is the limit of detection of the method (μg/kg). The LOD must meet the performance criteria set in point C.3.3.1 for the concentration of interest.
C is the concentration of interest (μg/kg);
α is a numeric factor to be used depending on the value of C. The values to be used are given in Table 10.
Table 10
Numeric values to be used for α as constant in formula set out in this point, depending on the concentration of interest
C (μg/kg)
α
≤ 50
0,2
51-500
0,18
501- 1 000
0,15
1 001 - 10 000
0,12
> 10 000
0,1
The analyst shall note the ‘Report on the relationship between analytical results, measurement uncertainty, recovery factors and the provisions of EU food and feed legislation’.

---