NoteWhen an automatic titration system is used, titration begins immediately after distillation starts and the 1 % boric acid solution (3.18) is used.
Where a fully automatic distillation unit is employed, the automatic titration of the ammonia can also be carried out with end-point detection using a potentiometric pH system.
In this case an automatic titrator, with a pH-meter is used. The pH-meter shall be calibrated properly in the range of pH 4 to pH 7 following normal laboratory pH-calibration procedures.
The pH end-point of the titration is reached at pH 4,6, being the steepest point in the titration curve (inflection point).

---