{
  "url": "https://openlegislation.uk/eu-regulation-2009-152/annex-III/division-P/division-1/division-5/division-5.3",
  "json_url": "https://openlegislation.uk/eu-regulation-2009-152/annex-III/division-P/division-1/division-5/division-5.3.json",
  "legislation": "eur/2009/152",
  "legislation_title": "Commission Regulation (EC) No 152/2009 of 27 January 2009 laying down the methods of sampling and analysis for the official control of feed (Text with EEA relevance)",
  "legislation_url": "https://openlegislation.uk/eu-regulation-2009-152",
  "path": "annex/III/division/P/division/1/division/5/division/5.3",
  "kind": "division",
  "number": "5.3",
  "heading": "Calibration curve",
  "text": "From the standard solution (3.7) prepare solutions containing respectively 5, 10, 20, 30 and 40 μg of phosphorus per ml. Take 10 ml of each of these solutions and add thereto 10 ml of molybdovanadate reagent (3.6). Homogenise and leave to stand for at least 10 minutes at 20 oC. Measure the optical density as indicated in 5.2. Trace the calibration curve by plotting the optical densities against the corresponding quantities of phosphorus. For concentrations between 0 and 40 μg/ml, the curve will be linear.",
  "extent": "E+W+S+N.I.",
  "status": null,
  "version": "current",
  "annotations": [],
  "outstanding_effects": [],
  "provenance": {
    "source": "legislation.gov.uk",
    "source_url": "https://www.legislation.gov.uk/eur/2009/152/annex/III/division/P/division/1/division/5/division/5.3",
    "retrieved_at": "2026-11-01T14:51:38.610873+00:00"
  },
  "licence": "Open Government Licence v3.0",
  "licence_url": "https://www.nationalarchives.gov.uk/doc/open-government-licence/version/3/",
  "attribution": "Contains public sector information licensed under the Open Government Licence v3.0. Source: legislation.gov.uk, The National Archives."
}