# Commission Regulation (EC) No 152/2009 of 27 January 2009 laying down the methods of sampling and analysis for the official control of feed (Text with EEA relevance), Division 5.5: Determination by HPLC

Extent: E+W+S+N.I.

Vitamin E is separated on a C18 reversed phase column (4.5.1) and the concentration is measured using a fluorescence detector (excitation: 295 nm, emission: 330 nm) or a UV detector (292 nm) (4.5.2).
Inject an aliquot portion (e.g. 20 μl) of the methanolic solution obtained in 5.4 and elute with the mobile phase (3.8). Calculate the mean peak heights (areas) of several injections of the same sample solution and the mean peak heights (areas) of several injections of the calibration solutions (5.6.2).

---
Source: https://www.legislation.gov.uk/eur/2009/152/annex/IV/division/B/division/5/division/5.5 (the official version)