The following conditions are offered for guidance; other conditions may be used provided that they give equivalent results.

Liquid chromatographic column (4.5.1):250 mm × 4 mm, C18, 5 or 10 μm packing, or equivalent
Mobile phase (3.8):Mixture of methanol (3.3) and water e.g. 980 + 20 (v + v).
Flow rate:1-2 ml/min.
Detector (4.5.2)

Fluorescence detector

(excitation: 295 nm/emission: 330 nm) or UV detector (292 nm)