OpenLegislation

Division 5.6.2.3 UV standardisation of the DL-α-tocopherol stock solution (3.11.1)

United Kingdom

Original text. Later amendments are not shown.

Dilute 2,0 ml of the DL-α-tocopherol stock solution (3.11.1) to 25,0 ml with ethanol and measure the UV spectrum of this solution against ethanol (3.1) in the spectrophotometer (4.6) between 250 nm and 320 nm. The absorption maximum shall be at 292 nm:

E1 %1 cm = 75,8 at 292 nm in ethanol

At this dilution an extinction value of 0,6 must be obtained.