Division 5.1.2.1 Preparation of calibration solutions

United Kingdom

Original text. Later amendments are not shown.

For each of the elements to be determined, prepare from the working standard solutions given in points 3.7.1, 3.8.1, 3.9.1 and 3.10.1 a range of calibration solutions, each calibration solution having an HCl concentration of about 0,5 mol/litre (and (in the cases of iron, manganese and zinc) a lanthanum chloride concentration equivalent to 0,1 % La (w/v).

The trace element concentrations selected must lie within the range of sensitivity of the spectrophotometer used. The tables below show, by way of example, the compositions of typical ranges of calibration solutions; depending, however, on the type and sensitivity of spectrophotometer used it may be necessary to select other concentrations.

  1. Iron
  2. Copper
  3. Manganese
  4. Zinc

Notes on Division 5

Footnotes

f1

Other methods of digestion may be used provided they have been demonstrated to have similar results (such as microwave pressure digestion).