# Commission Regulation (EC) No 152/2009 of 27 January 2009 laying down the methods of sampling and analysis for the official control of feed (Text with EEA relevance), Division 5.4.1: Parameters

Extent: E+W+S+N.I.

The following conditions are offered for guidance, other conditions may be used provided they yield equivalent results.
Liquid chromatographic column (4.4.1)
HPLC Mobile phase (3.21)
Flow rate: 1,5 to 2 ml/min.
Detection wavelength: 243 nm
Injection volume: 40 to 100 μl.
Check the stability of the chromatographic system, injecting the calibration solution (3.6.2) containing 3,0 μg/ml several times, until constant peak heights (or areas) and retention times are achieved.

---