Use of the lower band of the 95 % prediction interval at the decision limit of the confirmatory method:

Cut-off value=BEQDL−sy,x×t α,f=m−21∕n+1∕m+xi−x–2∕Qxx

with:

BEQ DL
  • BEQ corresponding to the decision limit of the confirmatory method, being the maximum level taking into account the expanded measurement uncertainty

  • s y,x
  • residual standard deviation

  • t α,f = m – 2
  • student factor (α = 5 %, f = degrees of freedom, single-sided)

  • m
  • total number of calibration points (index j)

  • n
  • number of repetitions on each level

  • x i
  • sample concentration (in TEQ) of calibration point i determined by a confirmatory method

  • x–
  • mean of the concentrations (in TEQ) of all calibration samples

  • Qxx=∑j=1mxi−x–2 square sum parameter, i = index for calibration point i