OpenLegislation

Division 3.7.1 Amprolium stock standard solution, 500 μg/ml

United Kingdom

Original text. Later amendments are not shown.

Weigh to the nearest 0,1 mg, 50 mg of amprolium (3.7) in a 100 ml graduated flask, dissolve in 80 ml methanol (3.1) and place the flask for 10 min. in an ultrasonic bath (4.4). After ultrasonic treatment bring the solution to room temperature, make up to the mark with water and mix. At a temperature of ≤ 4 oC the solution is stable for 1 month.