# Commission Regulation (EC) No 761/2009 of 23 July 2009 amending, for the purpose of its adaptation to technical progress, Regulation (EC) No 440/2008 laying down test methods pursuant to Regulation (EC) No 1907/2006 of the European Parliament and of the Council on the Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH) (Text with EEA relevance), Division 1.5.1.3: Procedure

Extent: E+W+S+N.I.

The substance is placed in the boiling section. Problems may be encountered with non-powder solids but these can sometimes be solved by heating the cooling jacket. The apparatus is sealed at the flange and the substance degassed. Frothing substances cannot be measured using this method.
The lowest desired pressure is then set and the heating is switched on. At the same time, the temperature sensor is connected to a recorder.
Equilibrium is reached when a constant boiling temperature is recorded at constant pressure. Particular care must be taken to avoid bumping during boiling. In addition, complete condensation must occur on the cooler. When determining the vapour pressure of low melting solids, care should be taken to prevent the condenser from blocking.
After recording this equilibrium point, a higher pressure is set. The process is continued in this manner until 105 Pa has been reached (approximately 5 to 10 measuring points in all). As a check, equilibrium points must be repeated at decreasing pressures.

---
Source: https://www.legislation.gov.uk/eur/2009/761/annex/I/division/1/division/1.5/division/1.5.1/division/1.5.1.3 (the official version)
Page: https://openlegislation.uk/eu-regulation-2009-761/annex-I/division-1/division-1.5/division-1.5.1/division-1.5.1.3