OpenLegislation

Division 1.5.3.3.2 Operation

United Kingdom

Original text. Later amendments are not shown.

The SEM should be operated at 10 000 magnificationf1 using conditions that give good resolution with an acceptable image at slow scan rates of, for example, 5 seconds per frame. Although the operational requirements of different SEMs may vary, generally to obtain the best visibility and resolution, with relatively low atomic weight materials, accelerating voltages of 5-10 keV should be used with a small spot size setting and short working distance. As a linear traverse is being conducted, then a 0° tilt should be used to minimise re-focussing or, if the SEM has a eucentric stage, the eucentric working distance should be used. Lower magnification may be used if the material does not contain small (diameter) fibres and the fibre diameters are large (> 5 μm).

Annotations

Footnotes

f1

For 3 μm fibres, see previous note.