Division Test protocol:X1
United Kingdom
Wells 1A and 1B are a blank control consisting of BTV antigen and conjugate. This may be used to blank the ELISA reader.
Columns 1 and 2, rows G and H are the monoclonal antibody control and contain BTV antigen, monoclonal antibody and conjugate. These wells represent maximum colour. The mean of the optical density readings from this control represents the 0 % inhibition value.
Columns 1 and 2, rows C-D-E-F. These wells contain BTV antigen, BTV strong and weak positive antiserum respectively, Mab and conjugate.
Wells 2A and 2B are the negative controls, which contain BTV antigen, BTV negative antiserum, Mab and conjugate.
For large-scale serological surveys and rapid screening, sera may be tested at a single dilution of 1:5 (Appendix 1). Alternatively, 10 sera may be tested over a dilution range from 1:5 to 1:640 (Appendix 2). This will give some indication of the titre of antibody in the test sera.