Division 1.2.7 Tissues or derivatives of animal origin coming into contact with the patients and users
United Kingdom
The manufacturer must consider:
- (a)
the maximum quantity of animal tissues or derivatives coming into contact with the patient or user when using a single medical device;
- (b)
the contact area: its surface, type (e.g. skin, mucous tissue, brain) and condition (e.g. healthy or damaged);
- (c)
the type of the tissues or derivatives coming into contact with the patients or users;
- (d)
the period of time the device is intended to remain in contact with the body (including bioresorption effect); and
- (e)
the number of medical devices that could be used in a given procedure or, if possible, over the lifetime of a patient or user.