OpenLegislation

Annex II

United Kingdom

Original text. Later amendments are not shown.

‘4.2.

General requirements

Confirmatory methods of analysis used for food control purposes shall comply with the provisions of items 1 and 2 of Annex III to Regulation (EC) No 882/2004.

4.3.

Specific requirements

4.3.1.

Specific requirements for confirmatory methods

4.3.1.1.

Performance criteria

It is recommended that fully validated confirmatory methods (i.e. methods validated by collaborative trials for relevant matrices) are used where appropriate and available. Other suitable validated confirmatory methods (e.g. methods validated in-house on relevant matrices belonging to the commodity group of interest) may also be used provided they fulfil the performance criteria set out in the following tables.

Where possible, the validation of in-house validated methods shall include a certified reference material.

  1. (a)

    Performance criteria for aflatoxins

    CriterionConcentration RangeRecommended ValueMaximum permitted Value
    BlanksAllNegligible—
    Recovery — Aflatoxin M10,01-0,05 mg/kg60 to 120 %
    > 0,05 mg/kg70 to 110 %
    Recovery-Aflatoxins B1, B2, G1, G2 < 1,0 mg/kg50 to 120 %
    1-10 mg/kg70 to 110 %
    > 10 mg/kg80 to 110 %
    Reproducibility RSDR AllAs derived from Horwitz Equation (*)(**)2 × value derived from Horwitz Equation (*)(**)
    Repeatability RSDr may be calculated as 0,66 times Reproducibility RSDR at the concentration of interest.

    Note:

    • Values to apply to both B1 and sum of B1 + B2 + G1 + G2

    • If sum of individual aflatoxins B1 + B2 + G1 + G2 are to be reported, then response of each to the analytical system must be either known or equivalent.

  2. (b)

    Performance criteria for ochratoxin A

    Levelμg/kgOchratoxin A
    RSDr %RSDR %Recovery %
    < 1≤ 40≤ 6050 to 120
    ≥ 1≤ 20≤ 3070 to 110
  3. (c)

    Performance criteria for patulin

    Levelμg/kgPatulin
    RSDr %RSDR %Recovery %
    < 20≤ 30≤ 4050 to 120
    20-50≤ 20≤ 3070 to 105
    > 50≤ 15≤ 2575 to 105
  4. (d)

    Performance criteria for deoxynivalenol

    Levelμg/kgDeoxynivalenol
    RSDr %RSDR %Recovery %
    > 100-≤ 500≤ 20≤ 4060 to 110
    > 500≤ 20≤ 4070 to 120
  5. (e)

    Performance criteria for zearalenone

    Levelμg/kgZearalenone
    RSDr %RSDR %Recovery %
    ≤ 50≤ 40≤ 5060 to 120
    > 50≤ 25≤ 4070 to 120
  6. (f)

    Performance criteria for Fumonisin B1 and B2 individually

    Levelμg/kgFumonisin B1 and B2 individually
    RSDr %RSDR %Recovery %
    ≤ 500≤ 30≤ 6060 to 120
    > 500≤ 20≤ 3070 to 110
  7. (g)

    Performance criteria for T-2 and HT-2 toxin individually

    Levelμg/kgT-2 and HT-2 toxin individually
    RSDr %RSDR %Recovery %
    15-250≤ 30≤ 5060 to 130
    > 250≤ 25≤ 4060 to 130
  8. (h)

    Performance criteria for citrinin

    Levelμg/kgCitrinin
    RSDr %Recommended RSDR %Maximum allowed RSDR %Recovery %
    All0,66 × RSDR As derived from Horwitz Equation (*)(**)2 × value derived from Horwitz Equation (*)(**)70 to 120
  9. (i)

    Notes to the performance criteria for the mycotoxins:

    • The detection limits of the methods used are not stated as the precision values given at the concentrations of interest.

    • The precision values are calculated from the Horwitz equation, in particular the original Horwitz equation (for concentrations 1,2 × 10–7 ≤ C ≤ 0,138) (*) and the modified Horwitz equation (for concentrations C < 1,2 × 10–7) (**).

      1. (*)

        Horwitz equation for concentrations 1,2 × 10–7 ≤ C ≤ 0,138:

        RSDR = 2(1-0.5logC)

        (ref: W. Horwitz, L.R. Kamps, K.W. Boyer, J.Assoc.Off.Analy.Chem.,1980, 63, 1344)

      2. (**)

        Modified Horwitz equation (*) for concentrations C < 1,2 × 10–7:

        RSDR = 22 %

        (ref: M. Thompson, Analyst, 2000, 125, p. 385-386)

        Where:

        • RSDR is the relative standard deviation calculated from results generated under reproducibility conditions [(sR/) × 100]

        • C is the concentration ratio (i.e. 1 = 100g/100g, 0,001 = 1 000 mg/kg)

        This is a generalised precision equation which has been found to be independent of analyte and matrix but solely dependent on concentration for most routine methods of analysis.

4.3.1.2.

“Fitness-for-purpose” approach

For in-house validated methods, as an alternative, a “fitness-for-purpose” approach (***) may be used to assess their suitability for official control. Methods suitable for official control must produce results with a standard measurement uncertainty (u) less than the maximum standard measurement uncertainty calculated using the formula below:

Uf = LOD ∕ 2 2 + α × C 2

where:

  • Uf is the maximum standard measurement uncertainty (μg/kg)

  • LOD is the limit of detection of the method (μg/kg)

  • α is a constant, numeric factor to be used depending on the value of C. The values to be used are set out in Table hereafter.

  • C is the concentration of interest (μg/kg)

If the analytical method provides results with uncertainty measurements less than the maximum standard uncertainty the method shall be considered being equally suitable to one which meets the performance criteria given in point 4.3.1.1.

Table

Numeric values to be used for α as constant in formula set out in this point, depending on the concentration of interest

C (μg/kg)α
≤ 500,2
51-5000,18
501-1 0000,15
1 001-10 0000,12
> 10 0000,1

(***)

Ref: M. Thompson and R. Wood, Accred. Qual. Assur., 2006, 10, p. 471-478.

4.3.2.

Specific requirements for semi-quantitative screening methods

4.3.2.1.

Scope

The scope applies to bioanalytical methods based on immuno-recognition or receptor binding (such as ELISA, dip-sticks, lateral flow devices, immuno-sensors) and physicochemical methods based on chromatography or direct detection by mass spectrometry (e.g. ambient MS). Other methods (e.g. thin layer chromatography) are not excluded provided the signals generated relate directly to the mycotoxins of interest and allow that the principle described hereunder is applicable.

The specific requirements apply to methods of which the result of the measurement is a numerical value, for example a (relative) response from a dip-stick reader, a signal from LC-MS, etc., and that normal statistics apply.

The requirements do not apply to methods that do not give numerical values (e.g. only a line that is present or absent), which require different validation approaches. Specific requirements for these methods are provided in point 4.3.3.

This document describes procedures for the validation of screening methods by means of an inter-laboratory validation, the verification of the performance of a method validated by means of an inter-laboratory exercise and the single-laboratory validation of a screening method.

4.3.2.2.

Terminology

Screening target concentration (STC): the concentration of interest for detection of the mycotoxin in a sample. When the aim is to test compliance with regulatory limits, the STC is equal to the applicable maximum level. For other purposes or in case no maximum level has been established, the STC is predefined by the laboratory.

Screening method: means method used for selection of those samples with levels of mycotoxins that exceed the screening target concentration (STC), with a given certainty. For the purpose of mycotoxin screening, a certainty of 95 % is considered fit-for-purpose. The result of the screening analysis is either “negative” or “suspect”. Screening methods shall allow a cost-effective high sample-throughput, thus increasing the chance to discover new incidents with high exposure and health risks to consumers. These methods shall be based on bio-analytical, LC-MS or HPLC methods. Results from samples exceeding the cut-off value shall be verified by a full re-analysis from the original sample by a confirmatory method.

“Negative sample” means the mycotoxin content in the sample is < STC with a certainty of 95 % (i.e. there is a 5 % chance that samples will be incorrectly reported as negative).

“False negative sample” means the mycotoxin content in the sample is > STC but it has been identified as negative.

“Suspect sample” (screen positive) means the sample exceeds the cut-off level (see below) and may contain the mycotoxin at a level higher than the STC. Any suspect result triggers a confirmatory analysis for unambiguous identification and quantification of the mycotoxin.

“False suspect sample” is a negative sample that has been identified as suspect.

“Confirmatory methods” means methods that provide full or complementary information enabling the mycotoxin to be identified and quantified unequivocally at the level of interest.

Cut-off level: the response, signal, or concentration, obtained with the screening method, above which the sample is classified as “suspect”. The cut-off is determined during the validation and takes the variability of the measurement into account.

Negative control (blank matrix) sample: a sample known to be freef1 of the mycotoxin to be screened for, e.g. by previous determination using a confirmatory method of sufficient sensitivity. If no blank samples can be obtained, then material with the lowest obtainable level might be used as long as the level allows the conclusion that the screening method is fit for purpose.

Positive control sample: sample containing the mycotoxin at the screening target concentration, e.g. a certified reference material, a material of known content (e.g. test material of proficiency tests) or otherwise sufficiently characterised by a confirmatory method. In the absence of any of the above, a blend of samples with different levels of contamination or a spiked sample prepared within laboratory and sufficiently characterised can be used, provided it can be proven that the contamination level has been verified.

4.3.2.3.

Validation procedure

The aim of the validation is to demonstrate the fitness of purpose of the screening method. This is done by determination of the cut-off value and determination of the false negative and false suspect rate. In these two parameters performance characteristics such as sensitivity, selectivity, and precision are embedded.

Screening methods can be validated by inter-laboratory or by single laboratory validation. If inter-laboratory validation data is already available for a certain mycotoxin/matrix/STC combination, a verification of method performance is sufficient in a laboratory implementing the method.

4.3.2.3.1.

Initial validation by single laboratory validation

Mycotoxins:

The validation shall be performed for every individual mycotoxin in the scope. In case of bio-analytical methods that give a combined response for a certain mycotoxin group (e.g. aflatoxins B1, B2, G1 & G2; fumonisins B1 & B2), applicability must be demonstrated and limitations of the test mentioned in the scope of the method. Undesired cross-reactivity (e.g. DON-3-glycoside, 3- or 15-acetyl-DON for immuno-based methods for DON) is not considered to increase the false negative rate of the target mycotoxins, but may increase the false suspect rate. This unwanted increasing will be diminished by confirmatory analysis for unambiguous identification and quantification of the mycotoxins.

Matrices:

An initial validation should be performed for each commodity, or, when the method is known to be applicable to multiple commodities, for each commodity group. In the latter case, one representative and relevant commodity is selected from that group (see table A).

Sample set:

The minimum number of different samples required for validation is 20 homogeneous negative control samples and 20 homogeneous positive control samples that contain the mycotoxin at the STC, analysed under intermediate precision (RSDRi) conditions spread over 5 different days. Optionally, additional sets of 20 samples containing the mycotoxin at other levels can be added to the validation set to gain insight to what extent the method can distinguish between different mycotoxin concentrations.

Concentration:

For each STC to be used in routine application, a validation has to be performed.

4.3.2.3.2.

Initial validation through collaborative trials

Validation through collaborative trials shall be done in accordance with an internationally recognised protocol on collaborative trials (e.g. ISO 5725:1994 or the IUPAC International Harmonised Protocol) which requires inclusion of valid data from at least eight different laboratories. Other than that, the only difference compared to single laboratory validations is that the ≥ 20 samples per commodity/level can be evenly divided over the participating laboratories, with a minimum of two samples per laboratory.

4.3.2.4.

Determination of cut-off level and rate of false suspected results of blank samples

The (relative) responses for the negative control and positive control samples are taken as basis for the calculation of the required parameters.

Screening methods with a response proportional with the mycotoxin concentration

For screening methods with a response proportional with the mycotoxin concentration the following applies:

Cut-off = RSTC – t-value0,05 * SDSTC

RSTC =
  • mean response of the positive control samples (at STC)

  • t-value:
  • one tailed t-value for a rate of false negative results of 5 % (see table B)

  • SDSTC =
  • standard deviationScreening methods with a response inversely proportional with the mycotoxin concentration

  • Similarly, for screening methods with a response inversely proportional with the mycotoxin concentration, the cut-off is determined as:

    Cut-off = RSTC + t-value0,05 * SDSTC

    By using this specific t-value for establishing the cut-off value, the rate of false negative results is by default set at 5 %.

    Fitness for purpose assessment

    Results from the negative control samples are used to estimate the corresponding rate of false suspect results. The t-value is calculated corresponding to the event that a result of a negative control sample is above the cut off value, thus erroneously classified as suspect.

    t-value
  • (cut off – meanblank)/SDblankfor screening methods with a response proportional with the mycotoxin concentration

  • or

    t-value
  • (mean blank – cut off)/SDblankfor screening methods with a response inversely proportional with the mycotoxin concentration

  • From the obtained t-value, based on the degrees of freedom calculated from the number of experiments, the probability of false suspect samples for a one tailed distribution can either be calculated (e.g.. spread sheet function “TDIST”) or taken from a table for t-distribution.

    The corresponding value of the one tailed t-distribution specifies the rate of false suspect results.

    This concept is described in detail with an example in Analytical and Bioanalytical Chemistry DOI 10.1007/s00216 -013-6922-1.

    4.3.2.5.

    Extension of the scope of the method

    4.3.2.5.1.

    Extension of scope to other mycotoxins:

    When new mycotoxins are added to the scope of an existing screening method, a full validation is required to demonstrate the suitability of the method.

    4.3.2.5.2.

    Extension to other commodities:

    If the screening method is known or expected to be applicable to other commodities, the validity to these other commodities shall be verified. As long as the new commodity belongs to a commodity group (see Table A) for which an initial validation has already been performed, a limited additional validation is sufficient. For this, a minimum of 10 homogeneous negative control and 10 homogeneous positive control (at STC) samples shall be analysed under intermediate precision conditions. The positive control samples shall all be above the cut-off value. In case this criterion is not met, a full validation is required.

    4.3.2.6.

    Verification of methods already validated through collaborative trials

    For screening methods that have already been successfully validated through a collaborative laboratory trial, the method performance shall be verified. For this a minimum of 6 negative control and 6 positive control (at STC) samples shall be analysed. The positive control samples shall all be above the cut-off value. In case this criterion is not met, the laboratory has to perform a root-cause analysis to identify why it cannot meet the specification as obtained in the collaborative trial. Only after taking corrective action it shall re-verify the method performance in its laboratory. In case the laboratory is not capable to verify the results from the collaborative trial, it will need to establish its own cut-off in a complete single laboratory validation.

    4.3.2.7.

    Continuous method verification/on-going method validation

    After initial validation, additional validation data are acquired by including at least two positive control samples in each batch of samples screened. One positive control sample is a known sample (e.g. one used during initial validation), the other is a different commodity from the same commodity group (in case only one commodity is analysed, a different sample of that commodity is used instead). Inclusion of a negative control sample is optional. The results obtained for the two positive control samples are added to the existing validation set.

    At least once a year the cut-off value is re-established and the validity of the method is re-assessed. The continuous method verification serves several purposes:

    • quality control for the batch of samples screened

    • providing information on robustness of the method at conditions in the laboratory that applies the method

    • justification of applicability of the method to different commodities

    • allowing to adjust cut-off values in case of gradual drifts over time.

    4.3.2.8.

    Validation report

    The validation report shall contain:

    • A statement on the STC

    • A statement on the obtained cut-off.

    Note: The cut-off must have the same number of significant figures as the STC. Numerical values used to calculate the cut-off need at least one more significant figure than the STC.

    • A statement on calculated false suspected rate

    • A statement on how the false suspected rate was generated.

    Note: The statement on the calculated false suspected rate indicates if the method is fit-for-purpose as it indicates the number of blank (or low level contamination) samples that will be subject to verification.

    Table A

    Commodity groups for the validation of screening methods

    If a buffer is used to stabilise the pH changes in the extraction step, then this commodity group can be merged into one commodity group “High water content”.

    “Difficult or unique commodities” should only be fully validated if they are frequently analysed. If they are only analysed occasionally, validation may be reduced to just checking the reporting levels using spiked blank extracts.

    Commodity groupsCommodity categoriesTypical representative commodities included in the category
    High water contentFruit JuicesApple juice, grape juice
    Alcoholic beveragesWine, beer, cider
    Root and tuber vegetablesFresh ginger
    Cereal or fruit based pureesPurees intended for infants and small children
    High oil contentTree nutsWalnut, hazelnut, chestnut
    Oil seeds and products thereofOilseed rape, sunflower, cotton-seed, soybeans, peanuts, sesame etc.
    Oily fruits and products thereofOils and pastes (e.g. peanut butter, tahina)
    High starch and/or protein content and low water and fat contentCereal grain and products thereofWheat, rye, barley, maize, rice, oats Wholemeal bread, white bread, crackers, breakfast cereals, pasta
    Dietary productsDried powders for the preparation of food for infants and small children
    High acid content and high water contentf2 Citrus products
    “Difficult or unique commodities”f3

    Cocoa beans and products thereof, copra and products thereof,

    coffee, tea

    Spices, liquorice

    High sugar low water contentDried fruitsFigs, raisins, currants, sultanas
    Milk and milk productsMilkCow, goat and buffalo milk
    CheeseCow, goat cheese
    Dairy products (e.g. milk powder)Yogurt, cream

    Table B

    One tailed t-value for a false negative rate of 5 %

    Degrees of FreedomNumber of replicatest-value (5 %)
    10111,812
    11121,796
    12131,782
    13141,771
    14151,761
    15161,753
    16171,746
    17181,74
    18191,734
    19201,729
    20211,725
    21221,721
    22231,717
    23241,714
    24251,711
    25261,708
    26271,706
    27281,703
    28291,701
    29301,699
    30311,697
    40411,684
    60611,671
    1201211,658
    ∞∞1,645

    4.3.3.

    Requirements for qualitative screening methods (methods that do not give numerical values)

    The development of validation guidelines for binary test methods is currently subject of various standardization bodies (e.g. AOAC, ISO). Very recently AOAC has drafted a guideline on this matter. This document can be regarded as the current state of the art in its field. Therefore methods that give binary results (e.g. visual inspection of dip-stick tests) should be validated according to this guideline

    http://www.aoac.org/imis15_prod/AOAC_Docs/ISPAM/Qual_Chem_Guideline_Final_Approved_031412.pdf

    4.4.

    Estimation of measurement uncertainty, recovery calculation and reporting of resultsf4

    4.4.1.

    Confirmatory methods

    The analytical result must be reported as follows:

    1. (a)

      Corrected for recovery, the level of recovery being indicated. The correction for recovery is not necessary in case the recovery rate is between 90-110 %.

    2. (b)

      As x +/– U whereby x is the analytical result and U is the expanded measurement uncertainty, using a coverage factor of 2 which gives a level of confidence of approximately 95 %.

    For food of animal origin, the taking into account of the measurement uncertainty can also be done by establishing the decision limit (CCα) in accordance with Commission Decision 2002/657/ECf5 (point 3.1.2.5 of Annex I — the case of substances with established permitted limit).

    However if the result of the analysis is significantly (> 50 %) lower than the maximum level or much higher than the maximum level (i.e. more than 5 times the maximum level), and on the condition that the appropriate quality procedures are applied and the analysis serves only the purpose of checking compliance with legal provisions, the analytical result might be reported without correction for recovery and the reporting of the recovery rate and measurement uncertainty might be omitted in these cases.

    The present interpretation rules of the analytical result in view of acceptance or rejection of the lot apply to the analytical result obtained on the sample for official control. In case of analysis for defence or referee purposes, the national rules apply.

    4.4.2.

    Screening methods

    The result of the screening shall be expressed as compliant or suspected to be non-compliant.

    “Suspected to be non-compliant” means the sample exceeds the cut-off level and may contain the mycotoxin at a level higher than the STC. Any suspect result triggers a confirmatory analysis for unambiguous identification and quantification of the mycotoxin.

    “Compliant” means that the mycotoxin content in the sample is < STC with a certainty of 95 % (i.e. there is a 5 % chance that samples will be incorrectly reported as negative). The analytical result is reported as “< level of STC” with the level of STC specified.’

    Annotations

    Footnotes

    f1

    Samples are considered free of analyte if the amount present in the sample does not exceed more than 1/5th of the STC. If the level can be quantified with a confirmatory method, the level must be taken into consideration for the validation assessment.

    f2

    If a buffer is used to stabilise the pH changes in the extraction step, then this commodity group can be merged into one commodity group “High water content”.

    f3

    “Difficult or unique commodities” should only be fully validated if they are frequently analysed. If they are only analysed occasionally, validation may be reduced to just checking the reporting levels using spiked blank extracts.

    f4

    More details on procedures for the estimation of measurement uncertainty and on procedures for assessing recovery can be found in the report “Report on the relationship between analytical results, measurement uncertainty, recovery factors and the provisions of EU food and feed legislation” — http://ec.europa.eu/food/food/chemicalsafety/contaminants/report-sampling_analysis_2004_en.pdf

    f5

    Commission Decision 2002/657/EC of 14 August 2002 implementing Council Directive 96/23/EC concerning the performance of analytical methods and the interpretation of results (OJ L 221, 17.8.2002, p. 8).