HPLC equipment, consisting of:
6.11.1. Pump
6.11.2. Injector, hand or automatic, with a 15 to 30 μl capacity
6.11.3. Two TSK 2 000 -SW columns in series (length 30 cm, internal diameter 0,75 cm) or equivalent columns (e.g. single TSK 2 000 -SWxl, single Agilent Technologies Zorbax GF 250) and a precolumn (3 cm × 0,3 cm) packed with I 125 or material of equivalent effectiveness
6.11.4. Thermostatic column oven, set at 35 ± 1 °C
6.11.5. Variable wavelength UV detector, permitting measurements at 205 nm with a sensitivity of 0,008 Å
6.11.6. Integrator capable of valley-to-valley integration
Note: Working with columns kept at room temperature is possible, but their power of resolution is slightly lower. In that case, the temperature should vary by less than ± 5 °C in any one range of analyses.

---