HPLC equipment, consisting of:
6.11.1. Binary gradient pumping system
6.11.2. Injector, hand or automatic, with a 100 μl capacity
6.11.3. Agilent Technologies Zorbax 300 SB-C3 column (length 25 cm, 0,46 cm internal diameter) or an equivalent wide-pore silica based reversed-phase column
6.11.4. Thermostatic column oven, set at 35 ± 1 °C
6.11.5. Variable wavelength UV detector, permitting measurements at 210 nm (if necessary, a higher wavelength up to 220 nm may be used) with a sensitivity of 0,02 Å
6.11.6. Integrator capable of setting the integration to common baseline or valley-to-valley
Note : Operation of the column at room temperature is possible, provided that the room temperature does not fluctuate more than 1 °C, otherwise too much variation in the retention time of CMP A takes place.

---