Division 2 Identification and confirmation of analytes of interest:

United Kingdom

Original text. Later amendments are not shown.

  • Relative retention time in relation to internal standards or reference standards (acceptable deviation of +/– 0,25 %).

  • Gas chromatographic separation of the non-dioxin-like PCBs (from interfering substances, especially co-eluting PCBs, in particular if levels of samples are in the range of legal limits and non-compliance is to be confirmedf1).

  • For GC-MS techniques:

    • Monitoring of at least the following number of molecular ions or characteristic ions from the molecular cluster:

      • two specific ions for HRMS,

      • three specific ions for LRMS,

      • two specific precursor ions, each with one specific corresponding transition product ion for MS-MS.

    • Maximum permitted tolerances for abundance ratios for selected mass fragments:

      Relative deviation of abundance ratio of selected mass fragments from theoretical abundance or calibration standard for target ion (most abundant ion monitored) and qualifier ion(s): ± 15 %.

  • For GC-ECD:

    Confirmation of results exceeding the maximum level with two GC columns with stationary phases of different polarity.

Annotations

Footnotes

f1

Congeners often found to co-elute are, e.g. PCB 28/31, PCB 52/69 and PCB 138/163/164. For GC-MS also possible interferences from fragments of higher chlorinated congeners have to be considered.