Division 8.1.10.1.5 HC FID methane (CH4) response verification

United Kingdom

Original text. Later amendments are not shown.

If the value of RF CH4[THC-FID] obtained in accordance with point 8.1.10.1.4 is within ± 5,0 % of its most recent previously determined value, the HC FID passes the methane response verification.

  1. (a)

    It shall be first verified that the pressures and / or flow rates of FID fuel, burner air, and sample are each within ± 0,5 % of their most recent previously recorded values, as described in point 8.1.10.1.3. If these flow rates have to be adjusted, a new RF CH4[THC-FID] shall be determined as described in point 8.1.10.1.4. It should be verified that the value of RF CH4[THC-FID] determined is within the tolerance specified in this point 8.1.10.1.5;

  2. (b)

    If RF CH4[THC-FID] is not within the tolerance specified in this point 8.1.10.1.5, the FID response shall be re-optimized as described in point 8.1.10.1.3;

  3. (c)

    A new RF CH4[THC-FID] shall be determined as described in point 8.1.10.1.4. This new value of RF CH4[THC-FID] shall be used in the calculations for HC determination, in section 2 of Annex VII (mass based approach) or section 3 of Annex VII (molar based approach).