# Commission Delegated Regulation (EU) 2017/654 of 19 December 2016 supplementing Regulation (EU) 2016/1628 of the European Parliament and of the Council with regard to technical and general requirements relating to emission limits and type-approval for internal combustion engines for non-road mobile machinery, Division 8.1.10.2.4: Procedure

Extent: E+W+S+N.I.

FID O2 interference shall be determined as follows, noting that one or more gas dividers may be used to create reference gas concentrations that are required to perform this verification:
(a) Three span reference gases that meet the specifications set out in point 9.5.1 and contain C3H8 concentration shall be selected to span the analyzers before emissions testing. CH 4 span reference gases shall be selected for FIDs calibrated on CH 4 with a non-methane cutter. The three balance gas concentrations shall be selected such that the concentrations of O2 and N2 represent the minimum and maximum and intermediate O2 concentrations expected during testing. The requirement for using the average O2 concentration can be removed if the FID is calibrated with span gas balanced with the average expected oxygen concentration;
(b) It shall be confirmed that the FID analyzer meets all the specifications of point 8.1.10.1;
(c) The FID analyzer shall be started and operated as it would be before an emission test. Regardless of the FID burner's air source during testing, zero air shall be used as the FID burner's air source for this verification;
(d) The analyzer shall be set at zero;
(e) The analyzer shall be spanned using a span gas that is used during emissions testing;
(f) The zero response shall be checked by using the zero gas used during emission testing. It shall be proceeded to the next step if the mean zero response of 30 s of sampled data is within ± 0,5 % of the span reference value used in paragraph (e) of this point, otherwise the procedure shall be restarted at paragraph (d) of this point;
(g) The analyzer response shall be checked using the span gas that has the minimum concentration of O2 expected during testing. The mean response of 30 s of stabilized sample data shall be recorded as x O2minHC;
(h) The zero response of the FID analyzer shall be checked using the zero gas used during emission testing. The next step shall be performed if the mean zero response of 30 s of stabilized sample data is within ± 0,5 % of the span reference value used in paragraph (e) of this point, otherwise the procedure shall be restarted at paragraph (d) of this point;
(i) The analyzer response shall be checked using the span gas that has the average concentration of O2 expected during testing. The mean response of 30 s of stabilized sample data shall be recorded as x O2avgHC;
(j) The zero response of the FID analyzer shall be checked using the zero gas used during emission testing. The next step shall be performed if the mean zero response of 30 s of stabilized sample data is within ± 0,5 % of the span reference value used in paragraph (e) of this point, otherwise the procedure shall be restarted at paragraph (d) of this point;
(k) The analyzer response shall be checked using the span gas that has the maximum concentration of O2 expected during testing. The mean response of 30 s of stabilized sample data shall be recorded as x O2maxHC;
(l) The zero response of the FID analyzer shall be checked using the zero gas used during emission testing. The next step shall be performed if the mean zero response of 30 s of stabilized sample data is within ± 0,5 % of the span reference value used in paragraph (e) of this point, otherwise the procedure shall be restarted at paragraph (d) of this point;
(m) The % difference between x O2maxHC and its reference gas concentration shall be calculated. The percent difference between x O2avgHC and its reference gas concentration shall be calculated. The % difference between x O2minHC and its reference gas concentration shall be calculated. The maximum % difference of the three shall be determined. This is the O2 interference;
(n) If the O2 interference is within ± 3 %, the FID passes the O2 interference verification; otherwise one or more of the following need to be performed to address the deficiency:
(i) The verification shall be repeated to determine if a mistake was made during the procedure;
(ii) The zero and span gases for emission testing shall be selected that contain higher or lower O2 concentrations and the verification shall be repeated;
(iii) The FID burner air, fuel, and sample flow rates shall be adjusted. Note that if these flow rates are adjusted on a THC FID to meet the O2 interference verification, the RF CH4 shall be reset for the next RF CH4 verification. The O2 interference verification shall be repeated after adjustment and RF CH4 shall be determined;
(iv) The FID shall be repaired or replaced and the O2 interference verification shall be repeated.

---
Source: https://www.legislation.gov.uk/eur/2017/654/annex/VI/division/8/division/8.1/division/8.1.10/division/8.1.10.2/division/8.1.10.2.4 (the official version)