Division 8.1.11.4.4 Procedure

United Kingdom

Original text. Later amendments are not shown.

The following procedure shall be used to verify sample dryer performance:

  1. (a)

    Instrument setup. The analyzer and sample dryer manufacturers' start-up and operating instructions shall be followed. The analyzer and sample dryer shall be adjusted as needed to optimize performance;

  2. (b)

    Equipment setup and data collection.

    1. (i)

      The total NOx gas analyzer(s) shall be zeroed and spanned as it would be before emission testing;

    2. (ii)

      NO2 calibration gas (balance gas of dry air) that has an NO2 concentration that is near the maximum expected during testing shall be selected. Higher concentration may be used in accordance with the instrument manufacturer's recommendation and good engineering judgement in order to obtain accurate verification, if the expected NO2 concentration is lower than the minimum range for the verification specified by the instrument manufacturer;

    3. (iii)

      This calibration gas shall be overflowed at the gas sampling system's probe or overflow fitting. Time shall be allowed for stabilization of the total NOx response, accounting only for transport delays and instrument response;

    4. (iv)

      The mean of 30 s of recorded total NOx data shall be calculated and this value recorded as x NOxref;

    5. (v)

      The flowing the NO2 calibration gas shall be stopped;

    6. (vi)

      Next the sampling system shall be saturated by overflowing a dew point generator's output, set at a dew point of 323 K (50 °C), to the gas sampling system's probe or overflow fitting. The dew point generator's output shall be sampled through the sampling system and sample dryer for at least 10 minutes until the sample dryer is expected to be removing a constant rate of water;

    7. (vii)

      It shall be immediately switched back to overflowing the NO2 calibration gas used to establish x NOxref. It shall be allowed for stabilization of the total NOx response, accounting only for transport delays and instrument response. The mean of 30 s of recorded total NOx data shall be calculated and this value recorded as x NOxmeas;

    8. (viii)

      x NOxmeas shall be corrected to x NOxdry based upon the residual water vapour that passed through the sample dryer at the sample dryer's outlet temperature and pressure;

  3. (c)

    Performance evaluation. If xNOxdry is less than 95 % of xNOxref, the sample dryer shall be repaired or replaced.