Division 8.1.5.3 System requirements
United Kingdom
- (a)
The system response time shall be ≤ 10 s with a rise time of ≤ 5 s for all measured components (CO, NOx, 2 and HC) and all ranges used.
All data (concentration, fuel and air flows) have to be shifted by their measured response times before performing the emission calculations given in Annex VII.
- (b)
To demonstrate acceptable updating and recording with respect to the system's overall response, the system shall meet one of the following criteria:
- (i)
The product of the mean rise time and the frequency at which the system records an updated concentration shall be at least 5. In any case the mean rise time shall be no more than 10 s;
- (ii)
The frequency at which the system records the concentration shall be at least 2 Hz (see also Table 6.7).
- (i)