Division 8.1.8.6.1 Specifications for differential flow measurement
United Kingdom
For partial flow dilution systems to extract a proportional raw exhaust gas sample, the accuracy of the sample flow qm p is of special concern, if not measured directly, but determined by differential flow measurement as set out in equation (6-20):
| qm p = qm dew – qm dw | (6-20) |
Where:
is the sample mass flow rate of exhaust gas into partial flow dilution system
is the dilution air mass flow rate (on wet basis)
is the diluted exhaust gas mass flow rate on wet basis
In this case, the maximum error of the difference shall be such that the accuracy of qm p is within ± 5 % when the dilution ratio is less than 15. It can be calculated by taking root-mean-square of the errors of each instrument.
Acceptable accuracies of q mp can be obtained by either of the following methods:
- (a)
The absolute accuracies of qm dew and qm dw are ± 0,2 % which guarantees an accuracy of qm p of ≤ 5 % at a dilution ratio of 15. However, greater errors will occur at higher dilution ratios;
- (b)
Calibration of qm dw relative to qm dew is carried out such that the same accuracies for qm p as in (a) are obtained. For details see point 8.1.8.6.2;
- (c)
The accuracy of q mp is determined indirectly from the accuracy of the dilution ratio as determined by a tracer gas, e.g. CO2. Accuracies equivalent to method (a) for q mp are required;
- (d)
The absolute accuracy of qm dew and qm dw is within ± 2 % of full scale, the maximum error of the difference between qm dew and qm dw is within 0,2 % and the linearity error is within ± 0,2 % of the highest qm dew observed during the test.