Division 8.1.9.1.4 Procedure
United Kingdom
The interference verification shall be performed as follows:
- (a)
The CO2 NDIR analyzer shall be started, operated, zeroed, and spanned as it would be before an emission test;
- (b)
F1A humidified test gas shall be created by bubbling zero air that meets the specifications set out in point 9.5.1 through distilled water in a sealed vessel. If the sample is not passed through a dryer, control the vessel temperature to generate an H 2 O content in the test gas at least as high as the maximum expected during testing. If the sample is passed through a dryer during testing, control the vessel temperature to generate an H 2 O content in the test gas at least as high as the maximum expected at the outlet of the dryer, in accordance with point 9.3.2.3.1.1;
- (c)
The humidified test gas temperature shall be maintained at least 5 oK above its dew point downstream of the vessel;
- (d)
The humidified test gas shall be introduced into the sampling system. The humidified test gas may be introduced downstream of any sample dryer, if one is used during testing;
- (e)
The water mole fraction, x H2O, of the humidified test gas shall be measured, as close as possible to the inlet of the analyzer. For example, dew point, T dew, and absolute pressure p total, shall be measured to calculate x H2O;
- (f)
Good engineering judgment shall be used to prevent condensation in the transfer lines, fittings, or valves from the point where x H2O is measured to the analyzer;
- (g)
Time shall be allowed for the analyzer response to stabilize. Stabilization time shall include time to purge the transfer line and to account for analyzer response;
- (h)
While the analyzer measures the sample's concentration, 30 s of sampled data shall be recorded. The arithmetic mean of this data shall be calculated. The analyzer meets the interference verification if this value is within (0,0 ± 0,4) mmol/mol.