Division 2.3 Measurement method

United Kingdom

Original text. Later amendments are not shown.

The reference laboratory shall measure each alignment tyre four times and retain the three last results for further analysis, in accordance with paragraph 4 of Annex 6 to UNECE Regulation No 117 and under the conditions set out in paragraph 3 of Annex 6 to UNECE Regulation No 117.

The candidate laboratory shall measure each alignment tyre (n + 1) times, with n being specified in Section 5 of this Annex and retain the n last results for further analysis, in accordance with paragraph 4 of Annex 6 to UNECE Regulation No 117 and applying the conditions set out in paragraph 3 of Annex 6 to UNECE Regulation No 117.

Each time an alignment tyre is measured, the tyre/wheel assembly shall be removed from the machine and the entire test procedure referred to in paragraph 4 of Annex 6 to UNECE Regulation No 117 shall be followed again from the start.

The candidate or reference laboratory shall calculate:

  1. (a)

    the measured value of each alignment tyre for each measurement as specified in paragraphs 6.2 and 6.3 of Annex 6 to UNECE Regulation No 117 (i.e. corrected for a temperature of 25 °C and a drum diameter of 2 m);

  2. (b)

    the mean value of the three last measured values of each alignment tyre (in the case of reference laboratories) or the mean value of the n last measured values of each alignment tyre (in the case of candidate laboratories); and

  3. (c)

    the standard deviation (σm ) as follows:

    where:

    i
  4. is the counter from 1 to p for the alignment tyres;

  5. j
  6. is the counter from 2 to n + 1 for the n last repetitions of each measurement of a given alignment tyre;

  7. n + 1
  8. is the number of repetitions of tyre measurements (n + 1 = 4 for reference laboratories and n + 1 ≥ 4 for candidate laboratories);

  9. p
  10. is the number of alignment tyres (p ≥ 5).