# The Genetically Modified Organisms(Contained Use) Regulations 2000, Part II


Table 1a:
Containment Measures for Activities Involving Genetic Modification of Micro-organisms in Laboratories
Containment Measures
Containment Levels
1
2
3
4
NOTES
1. In the Table above, “isolation” means, in relation to a laboratory, separation of the laboratory from other areas in the same building, or being in a separate building.
2. Entry must be through an airlock which is a chamber isolated from the laboratory. The clean side of the airlock must be separated from the restricted side by changing or showering facilities and preferably by interlocking doors.
3. Where viruses are not retained by the HEPA filters, extra requirements will be necessary for extract air.
4. Where the autoclave is outside the laboratory in which the activity involving genetic modification of micro-organisms is being undertaken, but within the laboratory suite, there shall be validated procedures for the safe transfer of material into that autoclave, which provide a level of protection equivalent to that which would be achieved by having an autoclave in that laboratory.
1
Laboratory suite: isolation (Note 1)
not required
not required
required
required
2
Laboratory: sealable for fumigation
not required
not required
required
required
Equipment
3
Surfaces impervious to water, resistant to acids, alkalis, solvents, disinfectants and decontamination agents and easy to clean
required for bench
required for bench
required for bench and floor
required for bench, floor ceiling and walls
4
Entry to lab via airlock (Note 2)
not required
not required
required where and to extent the risk assessment shows it is required
required
5
Negative pressure relative to the pressure of the immediate surroundings
not required
required where and to extent the risk assessment shows it is required
required
required
6
Extract and input air from the laboratory shall be HEPA filtered
not required
not required
HEPA filters required for extract air
HEPA filters required for input and extract air (Note 3)
7
Microbiological safety cabinet/enclosure
not required
required where and to extent the risk assessment shows it is required
required, and all procedures with infective materials required to be contained within a cabinet/enclosure
Class III cabinet required
8
Autoclave
required on site
required in the building
required in the laboratory suite (Note 4)
double ended autoclave required in laboratory
System of work
9
Access restricted to authorised personnel only
not required
required
required
required (via airlock key procedure)
10
Specific measures to control aerosol dissemination
not required
required so as to minimise
required so as to prevent
required so as to prevent
11
Shower
not required
not required
required where and to extent the risk assessment shows it is required
required
12
Protective clothing
suitable protective clothing required
suitable protective clothing required
suitable protective clothing required; footwear required where and to extent the risk assessment shows it is required
complete change of clothing and footwear required before entry and exit
13
Gloves
not required
required where and to extent the risk assessment shows they are required
required
required
14
Efficient control of disease vectors (eg rodents and insects) which could disseminate GMMs
required where and to extent the risk assessment shows it is required
required
required
required
15
Specified disinfection procedures in place
required where and to extent the risk assessment shows they are required
required
required
required
Waste
16
Inactivation of GMMs in effluent from handwashing sinks and showers and similar effluents
not required
not required
required where and to extent the risk assessment shows it is required
required
17
Inactivation of GMMs in contaminated material and waste
required by validated means
required by validated means
required by validated means
required by validated means
Other measures
18
Laboratory to contain its own equipment
not required
not required
required, so far as is reasonably practicable
required
19
An observation window or alternative is to be present so that occupants can be seen
required where and to extent the risk assessment shows it is required
required where and to extent the risk assessment shows it is required
required
required
20
Safe storage of GMMs
required where and to extent the risk assessment shows it is required
required
required
secure storage required
21
Written records of staff training
not required
required where and to extent the risk assessment shows they are required
required
required
Table 1b:
Containment Measures for Activities Involving Genetic Modification of Micro-organisms in Plant Growth Facilities (to be read with Table 1a as indicated in paragraph 3)
Containment Measures
Containment Levels
Additional/modification
1
2
3
4
NOTE
1. A permanent structure refers to a fixed structure with walls, a roof and a floor. Where the permanent structure is a greenhouse, that structure shall also have a continuous waterproof covering and self-closing lockable outer doors, and be located on a site designed to prevent the entry of surface run-off water.
Building
1
Permanent structure (Note 1)
required where and to extent the risk assessment shows it is required
required
required
required
Modification
Equipment
2
Entry via a separate room with two interlocking doors
not required
required where and to extent the risk assessment shows it is required
required where and to extent the risk assessment shows it is required
required (via airlock key procedure)
Additional
3
Control of contaminated run-off water
required where and to extent the risk assessment shows it is required
required so as to prevent run-off
required so as to prevent run-off
required so as to prevent run-off
Additional
System of work
4
Effective control of disease vectors such as insects, rodents and arthropods which could disseminate GMMs
required
required
required
required
Additional
5
Effective control of pollen, seeds and other plant material which could disseminate GMMs
required where and to extent the risk assessment shows it is required
required so as to minimise dissemination
required so as to prevent dissemination
required so as to prevent dissemination
Additional
6
Procedures for transfer of living material between the plant growth facilities, protective structure and laboratory shall control dissemination of GMMs
required so as to minimise dissemination
required so as to prevent dissemination
required so as to prevent dissemination
required so as to prevent dissemination
Additional
Table 1c:
Containment Measures for Activities Involving Genetic Modification of Micro-organisms in Animal Units (to be read with Table 1a as indicated in paragraph 3)
Containment Measures
Containment Levels
Additional/modification
1
2
3
4
NOTES
1. In the Table above, “animal unit” means a building, or separate area within a building, containing an animal facility and other areas including changing rooms, showers, autoclaves and food storage areas.
2. In the Table above and in Note 1 above, “animal facility” means a facility normally used to house stock, breeding or experimental animals or one which is used for the performance of minor surgical procedures on animals.
3. In the Table above, “isolators” means transparent boxes where small animals are contained within or outside a cage; for large animals, isolated rooms may be more appropriate.
Facilities
1
Isolation of animal unit (Note 1)
required where and to extent the risk assessment shows it is required
required
required
required
Modification
2
Animal facilities (Note 2) separated by lockable doors
required where and to extent the risk assessment shows they are required
required
required
required
Additional
3
Animal facilities (cages, etc) designed to facilitate decontamination (waterproof and easily washable material)
required where and to extent the risk assessment shows they are required
required where and to extent the risk assesment shows they are required
required
required
Additional
4
Floor, walls and ceiling easily washable
required where and to extent the risk assessment shows they are required
required for floor
required for floor and walls
required for floor, walls and ceiling
Modification
5
Appropriate filters on isolators or isolated rooms (Note 3)
not required
required where and to extent the risk assessment shows they are required
required
required
Additional
6
Incinerator for disposal of animal carcasses
required to be accessible
required to be accessible
required to be accessible
required to be on site
Additional
7
Appropriate barriers at the room exit, and at drains or ventilation duct work
required
required
required
required
Additional
8
Animals kept in appropriate containment facilities, such as cages, pens, tanks or isolators
required where and to extent the risk assessment shows it is required
required where and to extent the risk assessment shows it is required
required where and to extent the risk assessment shows it is required
required where and to extent the risk assessment shows it is required
Additional
Table 2:
Containment Measures for Activities Involving Genetic Modification of Micro-organisms in Premises other than those referred to in Tables 1a, 1b and 1c
Containment Measures
Containment Levels
1
2
3
4
General
1
Viable micro-organisms shall be contained in a system which separates the process from the workplace and wider environment (closed system)
required where and to extent the risk assessment shows it is required
required
required
required
2
Closed systems located within a controlled area
not required
required where and to extent the risk assessment shows they are required
required
required and required to be purpose built
3
Control of exhaust gases from the closed system
not required
required so as to minimise release
required so as to prevent release
required so as to prevent release
4
Control of aerosols during sample collection, addition of material to a closed system or transfer of material to another closed system
required where and to extent the risk assessment shows it is required
required so as to minimise release
required so as to prevent release
required so as to prevent release
5
Inactivation of bulk culture fluids before removal from the closed system
required where and to extent the risk assessment shows it is required
required by validated means
required by validated means
required by validated means
6
Seals shall be designed so as to minimise or prevent release
not required
required so as to minimise release
required so as to prevent release
required so as to prevent release
7
The controlled area designed to contain spillage of the entire contents of the closed system
required where and to extent the risk assessment shows it is required
required where and to extent the risk assessment shows it is required
required
required
8
The controlled area sealable to permit fumigation
not required
required where and to extent the risk assessment shows it is required
required where and to extent the risk assessment shows it is required
required
9
Biohazard signs posted
required where and to extent the risk assessment shows it is required
required
required
required
Equipment
10
Entry via airlock
not required
not required
required where and to extent the risk assessment shows it is required
required
11
Surfaces resistant to water, acids, alkalis, solvents, disinfectants and decontamination agents and easy to clean
required for any bench
required for any bench
required for floor and any bench
required for bench, floor, ceiling and walls
12
Specific measures to adequately ventilate the controlled areas in order to minimise air contamination
required where and to extent the risk assessment shows they are required
required where and to extent the risk assessment shows they are required
required where and to extent the risk assessment shows they are required
required
13
The controlled area maintained at an air pressure negative to the immediate surroundings
not required
not required
required where and to extent the risk assessment shows it is required
required
14
Extract and input air from the controlled area shall be HEPA filtered
not required
not required
required for extract air, optional for input air
required for input and extract air
System of work
15
Access restricted to authorised personnel only
not required
required
required
required
16
Decontamination and washing facilities provided for personnel
required
required
required
required
17
Personnel shall shower before leaving the controlled area
not required
not required
required where and to extent the risk assessment shows it is required
required
18
Personnel shall wear protective clothing
work clothing required
work clothing required
required
complete change required before exit and entry
19
Written procedures and records of staff training
not required
not required
required
required
Waste
20
Inactivation of GMMs in effluent from handwashing sinks and showers or similar effluents
not required
not required
required where and to extent the risk assessment shows it is required
required
21
Inactivation of GMMs in contaminated material and waste including those in process effluent before final discharge
required by validated means
required by validated means
required by validated means
required by validated means

---