# The Materials and Articles in Contact with Food (Scotland) Regulations 2012, Paragraph 4


Performance criteria of the instrumental method of analysis
(a) The detection limit for lead and cadmium must be equal to or lower than— 0,1 mg/litre for lead, 0,01 mg/litre for cadmium.
The detection limit is defined as the concentration of the element in the 4 % acetic acid solution, as referred to in paragraph 3(a) which gives a signal equal to twice the background noise of the instrument.
(b) The limit of quantification for lead and cadmium must be equal to or lower than— 0,2 mg/litre for lead, 0,02 mg/litre for cadmium.
(c) Recovery - the recovery of lead and cadmium added to the 4 % acetic acid solution, as referred to in paragraph 3(a) must lie within 80-120 % of the added amount.
(d) Specificity - the instrumental method of analysis used must be free from matrix and spectral interferences.

---