# The Measuring Instruments Regulations 2016, Paragraph 2.2


For quantities less than two litres the MPE on indications is:
Table 3
Measured volume V
MPE
V < 0.1 L
4 × value in Table 2, applied to 0.1 L
0.1 L ≤ V < 0.2 L
4 × value in Table 2
0.2 L ≤ V < 0.4 L
2 × value in Table 2, applied to 0.4 L
0.4 L ≤ V < 1 L
2 × value in Table 2
1 L ≤ V < 2 L
Value in Table 2, applied to 2 L

---
Source: https://www.legislation.gov.uk/uksi/2016/1153/schedule/1F/paragraph/2.2 (the official version)